9. TROUBLESHOOTING
This section shows the typical phenomena that imply troubles.
Some phenomena may be solved by rechecking or in simple manner.
Contact the nearest our distributor for phenomena that rechecking or
countermeasures may not eliminate or improve. Never detach the
case to check the inside.
Phenomenon
Possible cause
Check & countermeasures
Power is not turned
on even by throwing
the POWER switch.
・
AC power cable is broken
or has poor contact.
・
Check connection or replace the cable.
・
Input fuse has blown.
・
Input voltage is out of the allowable range.
“----“ is displayed after
starting operation.
The unit is specified as a
slave unit.
Specify the unit as the master unit. See
Section
“5-5-10. Master/slave function”.
“OHP” is displayed
after starting operation.
The OHP function works.
Check the ambient temperature, air
intake and exhaust ports and dust filter.
Display
disappears
in short time after
starting operation.
The fan is stopping.
Check the fan.
No output is given
even when the
OUTPUT key is
pressed.
・
The set voltage is
“0V”.
・
See Section
“5-4-1. How to set voltage”.
See Section
“5-7-2. Constant-voltage
control with external voltage or resistance
”.
・
The set current is
“0A”.
(In the CC status)
・
See Section
“5-4-2. How to set current”.
See Section
“5-7-3. Constant-current
control with external voltage or resistance.
”
・
On/Off with the external
contacts is selected.
・
See Section
“5-7-4. Output On/Off with
external contacts
”.
・
The channel output key is
not selected or an incorrect
channel is selected for
output.
・
See Section
“5-4-5. How to output”.
・
Keys are locked.
・
See Section
“5-4-8. How to invalidate
on-panel operations
”.
“OVP” is displayed
when the OUTPUT
key is pressed.
・
The OVP value is too low.
・
See Section
“5-5-2. OVP/OCP function”.
・
The sensing wire(s) is/are
disconnected.
・
See Section
“5-6. Voltage Remote
Sensing
”.
・
The resistor is open in the
external resistance control
mode.
・
See Section
“5-7-2. Constant-voltage
control with external voltage or
resistance
”.
“OCP” is displayed
when the OUTPUT
key is pressed.
・
The OCP value is too low.
See Section
“5-5-2. OVP/OCP function”.
Cannot
raise
the
current up to the set
current value.
CP (constant-power)
operation is carried out,
or the CP value is too low.
See Section
“5-4-3. How to set power”.
Output is unstable.
Oscillating due to
influences of the load.
・
Twist the load or sensing wires, or
change the wire layout.
・
Connect a capacitor at the output end
of the power source or the load end.
Output voltage does
not fall.
The Hi-
Ω
function is
activated.
See Section
“5-5-3. Hi-
Ω
function
”.
Содержание PSF-1200L
Страница 5: ......
Страница 11: ...2 400W 2 Channel Type PSF 400L2 400W Type PSF 400L 800W Type PSF 800L 800W Type For Current Extention PSF 800LS...
Страница 18: ...9...
Страница 115: ...106 PSF 800LS Dust filter Dust filter Front grill Front grill Pull frontward while pressing here...
Страница 117: ...108 290 35 7 6 5 82 7 209 9 10 9 124 The handle is available by option Fig 11 2 Outside Dimensions of PSF 800LS...
Страница 121: ...7F Towa Fudosan Shin Yokohama Bldg 2 18 13 Shin Yokohama Kohoku ku Yokohama Kanagawa 222 0033 Japan http www texio co jp...