Chapter 9 – Maintenance
Page 82
TestEquity 101H Temperature/Humidity Chamber
Dehumidification is accomplished through Capillary Tube (10) and the copper lines which are
bonded to the MTVG. The F4T Controller cycles the dehumidify coil liquid-line solenoid valve
(SV2, 7/8) ON/OFF through SSR3 to control the amount of dehumidification by condensing
moisture from the air. The moisture is condensed within the vertical space inside the MTVG.
The hot gas regulator (5) is adjusted to keep the suction pressure at 8 PSIG when both SV1 and
SV2 are off. A suction line cooling thermostatic expansion valve (6) senses the suction line
temperature and injects liquid refrigerant to cool the suction return gas within safe limits.
Содержание 101H
Страница 2: ......
Страница 6: ......
Страница 12: ......
Страница 42: ......
Страница 66: ......
Страница 70: ......
Страница 86: ......
Страница 88: ......
Страница 89: ...Chapter 11 Drawings TestEquity 101H Temperature Humidity Chamber Page 89 Chapter 11 Drawings...
Страница 90: ......
Страница 91: ...Electrical Subpanel Component Location R1...