Chapter 9 – Maintenance
Page 74
TestEquity 101H Temperature/Humidity Chamber
How to clean the condenser
1.
Unplug the chamber from the power source.
2.
Remove the condenser grille from the front of the chamber.
3.
Clean the condenser fins with a vacuum cleaner.
NOTE:
You may need to clean the condenser more frequently if the chamber is in a dusty
environment. You may be able to clean the condenser less frequently if the chamber is in a very
clean environment.
How to inspect the electrical/refrigeration compartment
!
WARNING:
Wear safety goggles when inspecting the electrical/refrigeration compartment
to protect against a refrigerant line which could break.
1.
Unplug the power cord.
2.
Remove the top and rear covers.
3.
Inspect for signs of refrigeration tubing abrasion.
4.
Inspect for oil around refrigeration valves, fittings and joints. This may be a sign of leaks.
5.
Inspect for loose hardware and tighten as required.
6.
Check for loose wires and burned insulation near terminals.
7.
Inspect for signs of insect or rodent infestation. Yes, it does happen!