Chapter 4 – Using the F4T Front Panel
TestEquity 101H Temperature/Humidity Chamber
Page 31
Sample Ramp and Soak Profile
Below is a typical ramp and soak temperature profile. This should give you an idea of how to
create your own custom profiles.
If this profile included humidity, the humidity values would be added at the
Set Point 2
prompts.
Additionally,
Humidity
would need to be
On
for any step that required humidity to be
controlled.
1.
Start at 23°C. Stay at 23°C for 30 minutes.
2.
Ramp to 85°C in 1 hour.
3.
Soak at 85°C for 2 hours, 30 minutes.
4.
Ramp to -40°C in 2 hours.
5.
Soak at -40°C for 2 hours, 30 minutes.
6.
Ramp to 85°C in 2 hours.
7.
Jump back to step 3. Repeat cycle 9 times for a total of 10 cycles.
8.
Go to 23°C as fast as possible. Wait until the temperatures reaches within 3° before
advancing to the next step.
9.
End with the chamber turned OFF.
Entering the above profile from the F4T front panel:
•
From the
Main Page
press
Profile Actions
.
•
Then press
Create Profile
on the screen that pops up. This will take you to the
New Profile
Screen
.
•
In the
New Profile
screen, enter your desired profile name in the
Name
field. You have the
choice of entering a password in the
Password
field if desired. In the
Guaranteed Soak
Deviation 1
field, enter
3
(this will be used in step 9 above).
•
You will now see
Step 1
.
o
Change step type to
Instant Change
.
o
Enter
23
for
Target Set Point 1
.
o
Enter
30
for
Minutes
.
o
Scroll down to see
Power
. Press
Off
and the press
ON
at the popup. This will
automatically be selected as
ON
for all subsequently added steps.
o
Press
Done
.
•
Press
Add Step
. You will now see
Step 2
.
o
Change step type to
Ramp Time
.
o
Enter
85
for
Target Set Point 1
.
o
Enter
1
for
Hour
and
0
for
Minutes
.
o
Press
Done
.
•
Press
Add Step
. You will now see
Step 3
.
o
Change step type to
Soak
.
o
Enter
2
for
Hour
and
30
for
Minutes
.
o
Press
Done
.
Содержание 101H
Страница 2: ......
Страница 6: ......
Страница 12: ......
Страница 42: ......
Страница 66: ......
Страница 70: ......
Страница 86: ......
Страница 88: ......
Страница 89: ...Chapter 11 Drawings TestEquity 101H Temperature Humidity Chamber Page 89 Chapter 11 Drawings...
Страница 90: ......
Страница 91: ...Electrical Subpanel Component Location R1...