Chapter 6 – Communications
Page 58
TestEquity 101H Temperature/Humidity Chamber
RS-232 Parameters
This section describes how to configure the RS-232 Modbus Communications settings in the
F4T controller.
Changing RS-232 Parameters
1.
Press the
Menu
,
Settings
and
Network
buttons, in that order.
2.
Under
Communications Channels
press
Modbus
.
3.
Change desired settings.
4.
Recycle the power to put the new settings into effect.
RS-232 Parameters and Settings
The bracketed bold settings below represent the defaults as delivered from the factory:
•
Modbus Address
: [1]
-
Each device is given a unique address. Up to 247 devices can be on the same data
link.
•
Baud Rate
: [9600]
-
Must match the baud rate of your computer’s serial port and software.
•
Parity
: [None]
-
Must match the parity of your computer’s serial port and software.
•
Display Units
: °F, °C [configured by TestEquity for °C]
•
Modbus Word Order
: [High], Low
-
Modbus allows a user to select the word order of two 16-bit words in floating point
values.
•
Data Map
: [1], 2
-
Data Map, the user can switch Modbus registers from the comprehensive listing of
F4T registers.
-
Map 1 = F4T registers (default)
-
Map 2 = Limited set of F4 compatible registers (for compatibility with software
written for the F4)
Содержание 101H
Страница 2: ......
Страница 6: ......
Страница 12: ......
Страница 42: ......
Страница 66: ......
Страница 70: ......
Страница 86: ......
Страница 88: ......
Страница 89: ...Chapter 11 Drawings TestEquity 101H Temperature Humidity Chamber Page 89 Chapter 11 Drawings...
Страница 90: ......
Страница 91: ...Electrical Subpanel Component Location R1...