C-3
Close the lid
Time Delay 1.5 sec
Move the probe upward
Time Delay 900mS
Move the handler outward
Time Delay 1sec
7 A
7 A
Start the timer
Timer = Timer + 1m sec
Timer = 1.4 sec?
Finish_status SET, Error_status
SET
Lid_status SET
Handler_status SET
Probe_status SET
Send Error Message to PC
<ERR30>
1B
Is the probe downward?
Move the probe downward
Downward sensor ON?
Finish_status SET, Error_status
RESET
Lid_status RESET
Handler_status RESET
Probe_status RESET
Send Response Message to PC
<READY>
N
Y
Y
N
Y
N
Appendix C TC-5942B Operation Flow Chart for programming