Check List for Common Problems
This section provides a brief checklist of common failures.
1. Verify that the AC power cord is connected to the TC-5942B.
Make sure that the power cord is firmly plugged into the power module on the left side. Also
make sure that the power source for the TC-5942B is energized.
2. Verify that the TC-5942B power switch is depressed.
3. Verify the line voltage.
The line voltage to the power source should be 100 ~ 240 VAC.
4. Verify the compressed air and air cables.
The compressed air to the TC-5942B should be 5 ~ 7.5 bar.
5. Verify external cables provide good connections.
6. Verify shield box specifications if the shield box is functional.
NOTE:
First verify that the problem is not a result of an external connection before considering the
problem is with the shield box. Then isolate the failure to the shield box before troubleshooting or
repairing the TC -5942B. Verify the shield box performance is to specification while troubleshooting
the problem.
5-3 Check List for Common Problems