Adjust the potentiometer until the “to process” temperature on the display equals
the temperature of the ice bath.
7.
When the two temperatures are equal, the calibration procedure is complete.
8.
Stop the unit following the shutdown procedure described in the manual
and place all process probes back in the unit to return to normal operation.
Calibration
potentiometer