08354B DCN7877
Teledyne API T640 PM Mass Monitor
80
MESSAGE
DESCRIPTION
POSSIBLE SOLUTION(S)
Replace pump.
Check Ext Pump
Check the external pump and/or
bypass flow control valve if Valve PWM
> 85%
Check pneumatics for blockages.
Check pneumatics for leaks.
Check external pump.
Check flow calibration.
External pump, or bypass flow
control valve may need replacing.
Sample Temp Warning Sample Temperature Warning (>60°C)
Check ASC (is it latched ON?)
Ensure proper climate and ventilation
for instrument.
Box Temp Warning
Box Temperature Warning (>60°C)
Ensure proper climate and ventilation
for instrument.
AMB Press Slope
OOR
Ambient Pressure Calibration Slope is
out of range
Check calibration (make sure units
match calibration device).
Replace pressure sensor.
Call Tech Support.
Time Sync Failure
Unable to synchronize time with NTP
server
Check network and/or internet
connection
Internal Serial Failure
Internal serial communication error with
sensor or control board
Check connections and wires to
board(s). Call Tech Support.
FLOW PROBLEMS
If a flow auditing device indicates any problems with flow, check to ensure the
following:
•
all connections are seated tightly and evenly (no gaps)
•
sensor inlet/nozzle, ASC, and/or sample line are not clogged or blocked.
•
pump is running and is within its PWM range (35% - 80%).
After making any adjustments, run a flow calibration and recheck the flow rate. If
problems persist, please contact Technical Support.
CALIBRATION PROBLEMS
Call our Technical Support Department for technical assistance.