Teledyne API T640 PM Mass Monitor
29
Figure 2-17. 640x Inlet Tube and Slip Coupler Assembly
IMPORTANT
IMPACT ON READINGS OR DATA
Ensure that the sample line is vertically plumb to prevent any strain
on the instrument and to prevent pneumatic leaks.
8.
Lightly tighten the tripod cord grip and the roof flange cord grip around the
sample line to hold it in place.
9.
At the top of the sample line (on the roof), be sure the provided inlet locking
collar is positioned 3” from the top of the sample line. This ensures the end of
the inlet tube is in the proper position within the inlet when the inlet is put in
place.
10.
Ensure the locking collar set screw is tightened onto the inlet tube.
11.
Place the inlet on top of the sample line.
•
Make sure the inlet is secure, i.e., that the collar holding it is not sliding.
•
Proper installation should have the inlet 2-m above the roof.
T640:
Figure 2-18. T640 Inlet and Sample Line Assembly