Getting Started
Teledyne API - Model T200H/T200M Operation Manual
38
3.4.6. CONNECTING THE CONTROL INPUTS
Control Inputs are used to remotely activate the zero and span calibration modes. Locate
the 10-pin connector labeled CONTROL IN on the analyzer’s rear panel.
There are two methods for energizing the control inputs. The in5V available
from the pin labeled “+” is the most convenient method. However, if full isolation is
required, an external 5 VDC power supply should be used.
S
P
AN CAL
Z
E
RO
CAL
L
O
W
S
P
AN
CONTROL IN
Local Power Connections
External Power Connections
S
P
AN CAL
Z
E
RO
CAL
L
O
W
S
P
AN
CONTROL IN
-
+
5 VDC Power
Supply
A B C D E F U
+
A B C D E F U
+
RANG
E
HI
RANG
E
HI
Figure 3-10:
Control Input Connector
Table 3-7: Control Input Signals
INPUT #
STATUS DEFINITION
ON CONDITION
A
REMOTE ZERO CAL
The analyzer is placed in Zero Calibration mode. The mode field of
the display will read
ZERO CAL R
.
B
REMOTE SPAN CAL
The analyzer is placed in Span Calibration mode. The mode field of
the display will read
SPAN CAL R
.
C
REMOTE LO SPAN CAL
The analyzer is placed in low span calibration mode as part of
performing a low span (midpoint) calibration. The mode field of the
display will read
LO CAL R
.
D
REMOTE RANGE HI
The analyzer is placed into high range when configured for dual
ranges..
E
SPARE
F
SPARE
Digital Ground
The ground level from the analyzer’s internal DC power supplies
(same as chassis ground).
U
External Power input
Input pin for +5 VDC required to activate pins A - F.
+
5 VDC output
Internally generated 5V DC power. To activate inputs A - F, place a
jumper between this pin and the “U” pin. The maximum amperage
through this port is 300 mA (combined with the analog output supply,
if used).
07270D DCN7141
Содержание T200H
Страница 2: ......
Страница 4: ...Teledyne API Model T200H T200M Operation Manual ii This page intentionally left blank 07270D DCN7141 ...
Страница 8: ...Teledyne API Model T200H T200M Operation Manual vi This page intentionally left blank 07270D DCN7141 ...
Страница 10: ...Teledyne API Model T200H T200M Operation Manual viii This page intentionally left blank 07270D DCN7141 ...
Страница 12: ...Teledyne API Model T200H T200M Operation Manual x This page intentionally left blank 07270D DCN7141 ...
Страница 344: ...This page intentionally left blank B 2 07270D DCN7141 ...
Страница 353: ...APPENDIX D 07270D DCN7141 07270D DCN7141 D 1 D 1 ...