Teledyne API - Model T200H/T200M Operation Manual
Troubleshooting & Service
263
QUESTION
ANSWER
Can I use the DAS system in
place of a strip chart recorder or
data logger?
Yes. Section 4.7 describes the setup and operation of the DAS system in
detail.
How often do I need to change
the particulate filter?
Once per week or as needed. Table 6-1 contains a maintenance schedule
listing the most important, regular maintenance tasks. Highly polluted
sample air may require more frequent changes.
How long does the sample pump
last?
The sample pump should last one to two years and the pump head should
be replaced when necessary. Use the
RCEL
pressure indicator on the front
panel to see if the pump needs replacement.
If this value goes above 10 in-Hg-A, on average, the pump head needs to
be rebuilt.
Why does my RS-232 serial
connection not work?
There are several possible reasons:
•
The wrong cable, please use the provided or a generic “straight-
through” cable (do not use a “null-modem” type cable),
•
The DCE/DTE switch on the back of the analyzer is not set properly;
make sure that both green and red lights are on,
•
The baud rate of the analyzer’s COM port does not match that of the
serial port of your computer/data logger. See Section 7.5.12 more
trouble-shooting information.
7.9. TECHNICAL ASSISTANCE
If this manual and its trouble-shooting / repair sections do not solve your problems,
technical assistance may be obtained from:
Teledyne-API, Technical Support
9970 Carroll Canyon Road, San Diego, CA 92131
Phone: +1 858 657 9800 or 1-800 324 5190
Fax: +1 858 657 9816
Email: [email protected].
Before you contact Technical Support, fill out the problem report form in Appendix C,
which is also available online for electronic submission at http://www.teledyne-
api.com/forms/.
07270D DCN7141
Содержание T200H
Страница 2: ......
Страница 4: ...Teledyne API Model T200H T200M Operation Manual ii This page intentionally left blank 07270D DCN7141 ...
Страница 8: ...Teledyne API Model T200H T200M Operation Manual vi This page intentionally left blank 07270D DCN7141 ...
Страница 10: ...Teledyne API Model T200H T200M Operation Manual viii This page intentionally left blank 07270D DCN7141 ...
Страница 12: ...Teledyne API Model T200H T200M Operation Manual x This page intentionally left blank 07270D DCN7141 ...
Страница 344: ...This page intentionally left blank B 2 07270D DCN7141 ...
Страница 353: ...APPENDIX D 07270D DCN7141 07270D DCN7141 D 1 D 1 ...