Teledyne API - Model T200H/T200M Operation Manual
Calibration Procedures
191
STEP THREE:
Perform Zero/Span Calibration:
Press
ENTR
to changes
the
OFFSET
&
SLOPE
values for both the NO
and NO
x
measurements.
Press
EXIT
to leave the
calibration unchanged and
return to the previous
menu.
SAMPLE
GAS TO CAL:NOX
NOX
O2
ENTR
EXIT
SAMPLE
NOX STB= XXX.X PPM
NOX=XXX.X
< TST TST > CAL SETUP
M-P CAL
NOX STB= XXX.X PPM
NOX=XXX.X
<TST TST>
ZERO
CONC
EXIT
SAMPLE
RANGE TO CAL:LOW
LOW
HIGH
ENTR
EXIT
EXIT
at this point
returns to the
SAMPLE menu.
Press
ENTR
to changes
the
OFFSET
&
SLOPE
values for both the NO
and NO
x
measurements.
Press
EXIT
to leave the
calibration unchanged and
return to the previous
menu.
SAMPLE A1:NXCNC1=100PPM NOX=XXX.X
< TST
TST >
CAL SETUP
Set the Display to show
the
NOX STB
test
function.
This function calculates
the stability of the NO/NO
x
measurement
Toggle
TST>
button until ...
Allow zero gas to enter the sample port
at the rear of the analyzer.
Wait until
NOX STB
falls below
0.5 ppm.
This may take several
minutes.
SAMPLE
NOX STB= XXX.X PPM
NOX=XXX.X
< TST TST >
CAL
SETUP
M-P CAL
NOX STB= XXX.X PPM
NOX=X.XXX
<TST TST>
ENTR
CONC
EXIT
SAMPLE
GAS TO CAL:NOX
NOX
O2
ENTR
EXIT
M-P CAL
NOX STB= XXX.X PPM
NOX=X.XXX
<TST TST> ZERO
SPAN
CONC
EXIT
SAMPLE
RANGE TO CAL:LOW
LOW
HIGH
ENTR
EXIT
Allow span gas to enter the sample port
at the rear of the analyzer.
Wait until
NOX STB
falls below
0.5 ppm.
This may take several
minutes.
SAMPLE
NOX STB= XXX.X PPM
NOX=XXX.X
< TST TST >
CAL
SETUP
M-P CAL
NOX STB= XXX.X PPM
NOX=X.XXX
<TST TST>
ENTR
CONC
EXIT
M-P CAL
NOX STB= XXX.X PPM
NOX=X.XXX
<TST TST> ENTR
CONC
EXIT
The
SPAN
key now appears
during the transition from
zero to span.
You may see both keys.
If either the
ZERO
or
SPAN
buttons fail to appear see
Section 11 for
troubleshooting tips.
Analyzer continues to
cycle through NO
x
,
NO, and NO
2
measurements
throughout this
procedure.
07270D DCN7141
Содержание T200H
Страница 2: ......
Страница 4: ...Teledyne API Model T200H T200M Operation Manual ii This page intentionally left blank 07270D DCN7141 ...
Страница 8: ...Teledyne API Model T200H T200M Operation Manual vi This page intentionally left blank 07270D DCN7141 ...
Страница 10: ...Teledyne API Model T200H T200M Operation Manual viii This page intentionally left blank 07270D DCN7141 ...
Страница 12: ...Teledyne API Model T200H T200M Operation Manual x This page intentionally left blank 07270D DCN7141 ...
Страница 344: ...This page intentionally left blank B 2 07270D DCN7141 ...
Страница 353: ...APPENDIX D 07270D DCN7141 07270D DCN7141 D 1 D 1 ...