Principles of Operation
Teledyne API - Model T200H/T200M Operation Manual
296
Table 8-4: Thermocouple Configuration Jumper (JP5) Pin-Outs
TC INPUT
JUMPER PAIR
DESCRIPTION
FUNCTION
TC1
1 – 11
Gain Selector
Selects preamp gain factor for J or K TC
- IN = J TC gain factor
- OUT = K TC gain factor
2 – 12
Output Scale Selector
Selects preamp gain factor for J or K TC
- IN = 5 mV / °C
- OUT = 10 mV / °C
3 – 13
Type J Compensation
When present, sets Cold Junction
Compensation for J type Thermocouple
4 – 14
Type K Compensation
When present, sets Cold Junction
Compensation for K type Thermocouple
5 – 15
Termination Selector
Selects
between Isolated and grounded TC
- IN =
Isolate TC
- OUT =
Grounded TC
TC2
6
– 1
6
Gain Selector
Same as Pins 1 – 11 above.
7 – 17
Output Scale Selector
Same as Pins 2 – 12 above.
8 – 18
Type J Compensation
Same as Pins 3 – 13 above.
9 – 19
Type K Compensation
Same as Pins 4 – 14 above.
10 – 20
Termination Selector
Same as Pins 5 – 15 above.
Figure 8-21:
Thermocouple Configuration Jumper (JP5) Pin-Outs
07270D DCN7141
Содержание T200H
Страница 2: ......
Страница 4: ...Teledyne API Model T200H T200M Operation Manual ii This page intentionally left blank 07270D DCN7141 ...
Страница 8: ...Teledyne API Model T200H T200M Operation Manual vi This page intentionally left blank 07270D DCN7141 ...
Страница 10: ...Teledyne API Model T200H T200M Operation Manual viii This page intentionally left blank 07270D DCN7141 ...
Страница 12: ...Teledyne API Model T200H T200M Operation Manual x This page intentionally left blank 07270D DCN7141 ...
Страница 344: ...This page intentionally left blank B 2 07270D DCN7141 ...
Страница 353: ...APPENDIX D 07270D DCN7141 07270D DCN7141 D 1 D 1 ...