Teledyne API - Model T200H/T200M Operation Manual
Calibration Procedures
199
The following example sets sequence #2 to carry out a Zero-Span calibration every other
day beginning January 1
st
, 2016, starting at 2:00 p.m. (1400 hours per the 24-hour clock
format), and 30 minutes later each day thereafter, lasting a total of 30 minutes each time
(15 minutes for Zero and 15 minutes for Span).
Table 5-4: Example Auto-Cal Sequence
MODE AND ATTRIBUTE
VALUE
COMMENT
SEQUENCE
2
Define sequence #2
MODE
ZERO-HI
Select zero and span mode
TIMER ENABLE
ON
Enable the timer
STARTING DATE
01-JAN-16
Start on or after 01 January 2016
STARTING TIME
14:00
First sequence starts at 14:00 (24-hour clock format)
DELTA DAYS
2
Repeat this sequence every 2 days
DELTA TIME
00:30
Repeat sequence 30 minutes later each time
(every 2 days and 30 minutes)
DURATION
15.0
Each sequence step will last 15 minutes (total of 30 minutes when
using zero-span mode)
CALIBRATE
ON
The instrument will recalculate the slope and offset values for the
NO and NO
X
channel at the end of the AutoCal sequence.
Please note the following suggestions for programming the AutoCal feature:
•
The programmed
Starting Time
must be 5 minutes later than the real time clock.
•
Avoid setting two or more sequences at the same time of the day. Any new
sequence which is initiated from a timer, the COM ports, or the contact closures will
override any sequence in progress. Note that two sequences with different daily
increments may eventually overlap.
•
If an illogical entry is selected, (for example: Delta Days > 366) the ENTR button will
disappear from the display.
•
With
CALIBRATE
turned on, the state of the internal setup variables
DYN_SPAN
and
DYN_ZERO
is set to
ON
and the instrument will reset the slope
and offset values for the NO and NO
X
response each time the AutoCal program
runs. This continuous re-adjustment of calibration parameters can often mask
subtle fault conditions in the analyzer. It is recommended that, if
CALIBRATE
is
enabled, the analyzer’s test functions, slope and offset values be checked
frequently to assure high quality and accurate data from the instrument.
07270D DCN7141
Содержание T200H
Страница 2: ......
Страница 4: ...Teledyne API Model T200H T200M Operation Manual ii This page intentionally left blank 07270D DCN7141 ...
Страница 8: ...Teledyne API Model T200H T200M Operation Manual vi This page intentionally left blank 07270D DCN7141 ...
Страница 10: ...Teledyne API Model T200H T200M Operation Manual viii This page intentionally left blank 07270D DCN7141 ...
Страница 12: ...Teledyne API Model T200H T200M Operation Manual x This page intentionally left blank 07270D DCN7141 ...
Страница 344: ...This page intentionally left blank B 2 07270D DCN7141 ...
Страница 353: ...APPENDIX D 07270D DCN7141 07270D DCN7141 D 1 D 1 ...