Teledyne API - Model T200H/T200M Operation Manual
Getting Started
37
3.4.5.1. Converting Current Loop Analog Outputs to Standard Voltage Outputs.
To convert an output configured for current loop operation to the standard 0 to 5 VDC
output operation:
1. Turn off power to the analyzer.
2. If a recording device was connected to the output being modified, disconnect it.
3. Remove the top cover:
a. Remove the set screw located in the top, center of the rear panel
b. Remove the screws fastening the top cover to the unit (four per side).
c. Lift the cover straight up.
4. Disconnect the current loop option PCA from the appropriate connector on the
motherboard.
5. Place a shunt between the leftmost two pins of the connector (see Figure 3-9).
6. Reattach the top case to the analyzer.
The analyzer is now ready to have a voltage-sensing, recording device attached to that
output.
CAUTION – Avoid Warranty Invalidation
Printed Circuit Assemblies (PCAs) are sensitive to electro-static discharges too small to be
felt by the human nervous system. Failure to use Electro-Static Discharge (ESD) protection
when working with electronic assemblies will void the instrument warranty. Refer to the
manual, Fundamentals of ESD, PN 04786, which can be downloaded from our website at
under Help Center > Product Manuals in the Special Manuals
section.
07270D DCN7141
Содержание T200H
Страница 2: ......
Страница 4: ...Teledyne API Model T200H T200M Operation Manual ii This page intentionally left blank 07270D DCN7141 ...
Страница 8: ...Teledyne API Model T200H T200M Operation Manual vi This page intentionally left blank 07270D DCN7141 ...
Страница 10: ...Teledyne API Model T200H T200M Operation Manual viii This page intentionally left blank 07270D DCN7141 ...
Страница 12: ...Teledyne API Model T200H T200M Operation Manual x This page intentionally left blank 07270D DCN7141 ...
Страница 344: ...This page intentionally left blank B 2 07270D DCN7141 ...
Страница 353: ...APPENDIX D 07270D DCN7141 07270D DCN7141 D 1 D 1 ...