Operating Instructions
Teledyne API - Model T200H/T200M Operation Manual
148
To initiate an automatic calibration for an individual output channel, press:
From the
AIO CONFIGURATION SUBMENU
(in the SETUP DIAG menu)
DIAG AIO DATA_OUT_2 5V, NXCNC1, NOCAL
<SET
SET>
EDIT
EXIT
Continue pressing
SET>
until ...
DIAG AIO DATA_OUT_2 CALIBRATED:NO
<SET
SET>
CAL
EXIT
DIAG AIO DATA_OUT_2 RANGE: 5V
SET>
EDIT
EXIT
DIAG AIO
AUTO CALIBRATING DATA_OUT_2
DIAG AIO DATA_OUT_2 CALIBRATED: YES
<SET
SET> CAL
EXIT
DIAG
ANALOG I/O CONFIGURATION
PREV NEXT
ENTR
EXIT
Continue pressing
SET>
until you reach the
output to be configured
DIAG AIO AOUTS CALIBRATED: NO
SET>
CAL
EXIT
07270D DCN7141