Operating Instructions
Teledyne API - Model T200H/T200M Operation Manual
134
4.13.5.2. Analog Output Signal Type and Range Selection
To select an output signal type (DC Voltage or current) and level for one output channel
press:
DIAG AIO AOUTS CALIBRATED: NO
SET>
CAL
EXIT
From the
AIO CONFIGURATION SUBMENU
(in the SETUP DIAG menu)
DIAG
ANALOG I/O CONFIGURATION
PREV NEXT
ENTR
EXIT
Continue pressing
SET>
until you reach the
output to be configured
DIAG AIO DATA_OUT_3: 5V, NXCNC1, NOCAL
<SET
SET>
EDIT
EXIT
DIAG AIO DATA_OUT_3: RANGE: 5V
0.1V
1V
5V
10V CURR
ENTR EXIT
These keys set
the signal level
and type of the
selected
channel
Pressing
ENTR
records
the new setting and
returns to the previous
menu.
Pressing
EXIT
ignores the
new setting and returns to
the previous menu.
07270D DCN7141