
Getting Started
Teledyne API - Model T200H/T200M Operation Manual
32
3.4. ELECTRICAL CONNECTIONS
Note
To maintain compliance with EMC standards, it is required that the cable length
be no greater than 3 meters for all communication connections.
Refer to Figure 3-4 for the location of the rear panel electrical and pneumatic
connections.
3.4.1. POWER CONNECTION
Attach the power cord to the analyzer and plug it into a power outlet capable of carrying
at least 10 A current at your AC voltage and that it is equipped with a functioning earth
ground.
CAUTION
ENSURE THAT the voltage and frequency specifications on the rear panel Label of the
instrument ARE compatibLE with the local power before plugging the T200H/M into line
power.
Do not plug in the power cord if the voltage or frequency is incorrect.
WARNING – RISK OF ELECTRIC SHOCK
•
High voltages are present inside the instrument’s chassis.
•
Power connection must have functioning ground connection.
•
Do not defeat the ground wire on power plug.
•
Turn off analyzer power before disconnecting or connecting electrical
subassemblies.
•
Do not operate with cover off.
07270D DCN7141
Содержание T200H
Страница 2: ......
Страница 4: ...Teledyne API Model T200H T200M Operation Manual ii This page intentionally left blank 07270D DCN7141 ...
Страница 8: ...Teledyne API Model T200H T200M Operation Manual vi This page intentionally left blank 07270D DCN7141 ...
Страница 10: ...Teledyne API Model T200H T200M Operation Manual viii This page intentionally left blank 07270D DCN7141 ...
Страница 12: ...Teledyne API Model T200H T200M Operation Manual x This page intentionally left blank 07270D DCN7141 ...
Страница 344: ...This page intentionally left blank B 2 07270D DCN7141 ...
Страница 353: ...APPENDIX D 07270D DCN7141 07270D DCN7141 D 1 D 1 ...