Operating Instructions
Teledyne API - Model T200H/T200M Operation Manual
98
Example: If all passwords are enabled, the following menu button sequence would be
required to enter the SETUP menu:
SAMPLE A1:NXCNC1=100PPM NOX=XXX.X
< TST TST > CAL
SETUP
SAMPLE
ENTER SETUP PASS: 0
0 0 0
ENTR EXIT
prompts for
password
number
Example: this
password enables the
SETUP mode
SAMPLE
ENTER SETUP PASS: 0
8 1 8
ENTR EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE
EXIT
Press individual
buttons to set
numbers
Note that the instrument still prompts for a password when entering the VARS and
DIAG menus, even if passwords are disabled, but it displays the default password (818)
upon entering these menus. The user only has to press ENTR to access the password-
protected menus but does not have to enter the required number code.
07270D DCN7141