Operating Instructions
Teledyne API - Model T200H/T200M Operation Manual
150
To manually adjust the signal levels of an analog output channel, press:
From the
AIO CONFIGURATION SUBMENU
(in the SETUP DIAG menu)
DIAG
ANALOG I/O CONFIGURATION
PREV NEXT
ENTR
EXIT
Continue pressing
SET>
until you reach the
output to be configured
DIAG AIO DATA_OUT_2 5V, NXCNC1, NOCAL
<SET
SET>
EDIT
EXIT
Continue pressing
SET>
until ...
DIAG AIO DATA_OUT_2 CALIBRATED:NO
<SET
SET>
CAL
EXIT
DIAG AIO DATA_OUT_2 RANGE: 5V
SET>
EDIT
EXIT
DIAG AIO DATA_OUT_2 VOLT-Z: 0 mV
U100 UP10 UP DOWN DN10 D100 ENTR
EXIT
DIAG AIO DATA_OUT_2 VOLT-S: 4500 mV
U100 UP10 UP DOWN DN10 D100 ENTR EXIT
DIAG AIO DATA_OUT_2 CALIBRATED: YES
<SET
SET> CAL
EXIT
These menus
only appear if
AUTO-CAL
is
turned
OFF
These buttons increase / decrease
the analog output signal level (not the
value on the display)
by 100, 10 or 1 counts.
Continue adjustments until the
voltage measured at the output of the
analyzer and/or the input of the
recording device matches the value
in the upper right hand corner of the
display (within the voltage tolerances
listed in the preceding table
)
.
DIAG AIO AOUTS CALIBRATED: NO
SET>
CAL
EXIT
4.13.6.3. Manual Calibration of Analog Outputs Configured for Current Loop Ranges
The current loop output option (see Section 3.4.5) uses a small converter assembly to
change the DC voltage output by the standard voltage output to a current signal ranging
between 0-20 mA. Since the exact current increment per voltage count varies from
converter to converter and from instrument to instrument, analog outputs with this
option installed cannot be calibrated automatically and must be adjusted manually.
Adjusting the signal zero and full scale values of the current loop output is done in a
similar manner as manually adjusting analog outputs configured for voltage output
except that:
•
In this case calibration is performed with a current meter connected in series with
the output circuitry (See Figure 4-12).
•
Adjustments to the output are made using the front panel touchscreen, also in 100,
10 or 1 count increments, but the change in the voltage driving the converter
assembly is displayed on the front panel.
•
As before, adjustment of the output is performed until the current reading of the
meter reaches the desired point (e.g. 2 mA, 4 mA, 20 mA, etc.)
07270D DCN7141
Содержание T200H
Страница 2: ......
Страница 4: ...Teledyne API Model T200H T200M Operation Manual ii This page intentionally left blank 07270D DCN7141 ...
Страница 8: ...Teledyne API Model T200H T200M Operation Manual vi This page intentionally left blank 07270D DCN7141 ...
Страница 10: ...Teledyne API Model T200H T200M Operation Manual viii This page intentionally left blank 07270D DCN7141 ...
Страница 12: ...Teledyne API Model T200H T200M Operation Manual x This page intentionally left blank 07270D DCN7141 ...
Страница 344: ...This page intentionally left blank B 2 07270D DCN7141 ...
Страница 353: ...APPENDIX D 07270D DCN7141 07270D DCN7141 D 1 D 1 ...