LabMaster 10 Zi Oscilloscopes
80
LM10Zi-GSM-E Rev A
instances when a
Num Segments
miscount occurs for some
reason and the oscilloscope waits indefinitely for an
unforthcoming segment. During that time, no oscilloscope
functions are accessible. By means of a timeout value, however,
the acquisition will be completed, the waveform displayed, and
control of the oscilloscope returned to the user after the timeout
has elapsed.
6.
Under
Display Settings
, touch inside the
Display mode
control,
and select a sequence mode display from the pop-up menu.
NOTE
: Once a single acquisition has started, you can interrupt the
acquisition at any time by pressing the
S
INGLE
front panel button
a second time or by pressing the
S
TOP
front panel button. In this
case, the segments already acquired will be retained in memory.
7.
Touch the
S
INGLE
trigger front panel button.