LabMaster 10 Zi Oscilloscopes
156
LM10Zi-GSM-E Rev A
Pass/Fail Boolean Conditions
All True
All False
Any True
Any False
All Q1 to Q4
Or
All Q5 to Q8
Any Q1 to Q4
And
Any Q5 to Q8
P
ASS
/F
AIL
T
ESTING
S
ETUP
Initial Setup
1.
Touch
Analysis → Pass/Fail Setup...
on the menu bar.
2.
Touch the
Actions
tab.
3.
Touch the
Enable Actions
checkbox. This causes selected actions
to occur on your waveform's passing or failing a test.
4.
Touch the
Summary View
checkbox to enable a line of text (as
follows) showing a concise status of the last waveform and
keeping a running count of how many sweeps have passed.
5.
Touch inside the
Pass If
control, and select a Boolean condition
from the choices shown.
6.
If you want to set up the test to end after a finite number of
sweeps, touch the
Stop Test
checkbox. Then touch inside the
After
data entry control and provide a value using your preferred
input control method. Additional information on using the touch
screen controls can be found in the
Dialog Area
.
7.
Use the
Pass
or
Fail
button on the
If
section of
the dialog to set actions taking place when your waveform passes
or fails the test.
8.
The
Then
section of the dialog provides
stop test
,
sound alarm
,
print result
,
emit pulse
, or
save
(the waveform) actions. If you