Delay Setting constant in
- allows you to specify how Horizontal scale
changes are displayed. Choose either:
Time
- As the timebase changes, keep the horizontal offset
indicator stationary. With this selection, the trigger point could
move off the grid.
Div
(isions) - Have the horizontal offset indicator move with the
trigger point. With this selection, the trigger point remains on the
grid as you increase the timebase.
NOTE
: Regardless of whether you select Time or Div, the Delay shown in
the Timebase Setup dialog always indicates time.
Reset trigger counter before starting a new acquisition
- If available,
this selection clears the trigger counter each time the oscilloscope
issues an ARM acquisition command and applies any trigger Holdoff
conditions set on the Trigger dialog: