Getting Started Guide
LM10Zi-GSM-E Rev A
5
AC Power Source
100 to 240 VAC (±10%) at 50/60 Hz (± 10%).
Manual voltage selection is not required because the instrument
automatically adapts to line voltage.
Power Consumption
MCM-Zi Master Control Module
≤ 450 watts (450 VA) (depending on accessories installed – PC
port plug-ins etc.).
Acquisition Modules
25-36 GHz Models
- ≤ 1100 watts (1100 VA) for 25-36 GHz
LabMaster 10 Zi series models (depending on accessories
installed - probes, serial trigger option, etc.).
50-65 GHz Models
- ≤ 1150 watts (1150 VA) for 30-45 GHz
LabMaster 10 Zi series
DBI
models (depending on accessories
installed - probes, serial trigger option, etc.).
Power and Ground Connections
The MCM-Zi Master Control Module is provided with a 10A/250V rated
grounded cord set containing a molded three-terminal polarized plug
with a standard IEC-60320 (Type C13) connector. The connector mates
to a compatible power inlet on the Master Control Module for making
line voltage and safety ground connections.
The 10-xxZi Acquisition Module is provided with a 16A/250V or
15A/125V rated grounded cord set containing a molded three-terminal
polarized plug and a specific IEC-60320 (Type C19) connector. The
connector mates to a compatible power inlet on the Acquisition Module
for making line voltage and safety ground connections.
For both modules, power inlet ground terminal is connected directly to
the frame of the respective instrument. For adequate protection