LabMaster 10 Zi Oscilloscopes
128
LM10Zi-GSM-E Rev A
PLEASE NOTE THE FOLLOWING
:
The bottom cursor knob adjusts relative cursors. It does not work
with absolute cursor types.
Previously setup cursor positions are preserved for subsequent
displays. Changing cursor positions or measurement modes are
always available using touch screen or front panel control
methods as previously mentioned.
The Cursors Dialog
After using one of the above methods to access cursors, the
Cursors
dialog is displayed.
C
URSORS
O
N
Mark or unmark the
Cursors On
checkbox to turn cursors on and off as
desired.
C
URSOR
T
YPES
Select your
Cursor Type
from
Horizontal Abs
,
Horizontal Rel
,
Vertical
Abs
, and
Vertical Rel
options shown. Subsequent controls on the dialog
are shown, hidden, enabled, or disabled based on the
Cursor Type
selection made.
Horizontal Abs
- Provides a single, dashed, vertical line and cross-
hair marking for the specific cursor location. None of the
Show
controls are provided; however, the Position 1 control is enabled
on the right of the dialog.
Horizontal Rel
- Provides two, dashed, vertical lines and cross-
hair markings for the specific cursor locations. Since there are
two cursors involved, controls for
Show
along with
Track
, and
Position 1
and
2
are available on the dialog.