CombiFlash Torrent Installation Guide

Содержание CombiFlash Torrent

Страница 1: ...der computer control at user selected flow rates between 50 and 1000 mL min The Torrent operates up to 100 psi 6 9 bar and can use low or medium pressure Flash columns for either normal or reversed ph...

Страница 2: ...e control via Windows PC Software On screen method development and control with PeakTrak software Fraction collection Up to four fraction collectors fractionation valve or manual collection Solvent ma...

Страница 3: ...local regulations In the event of a solvent or chemical spill your organization should have a plan to deal with these mishaps In all cases use good laboratory practices and standard safety procedures...

Страница 4: ...ymbols Symboles de s curit Ce symbole signale l existence d instructions importantes relatives au produit dans ce manuel Ce symbole signale la pr sence d un danger d lectrocution Risque de pincement C...

Страница 5: ...mail IscoService teledyne com Advertencias y Precauciones Esta se al le advierte sobre la importancia de las instrucciones del manual que acompa an a este producto Esta se al alerta sobre la presencia...

Страница 6: ...CombiFlash Torrent Installation Guide...

Страница 7: ...rent is shipped in a single carton The optional equipment will be shipped in additional cartons Carefully unpack the shipment and inspect the contents WARNING The Torrent is heavy Use a two person lif...

Страница 8: ...be connected to a user supplied mast 2 5 Connect Power Ensure that the On Standby switch above the touch screen panel is in the Standby position Then use the supplied IEC power cord to connect the Tor...

Страница 9: ...ction collectors via an Ethernet network To connect one or the first of multiple Foxys connect a standard straight through network cable between the Ethernet ports on the Torrent and a Foxy The second...

Страница 10: ...ed Keep in mind that the valve wash waste is located on the Torrent s left side panel diverter valve waste will come from the fraction collector or fractionation valve usually positioned on the right...

Страница 11: ...isopropyl alcohol some of which may drain when the caps are removed The Torrent may receive solvent from User supplied solvent containers at atmospheric pressure Pressurized or gravity fed solvent de...

Страница 12: ...ional 1 8 turn 8 The Solvent Select Valve should be plumbed to the B Inlet of the Torrent Section 2 2 6 9 Locate the Sample Inlet line This assembly consists of two tubes fastened together a white sol...

Страница 13: ...ngs from the source to the A Inlet and B Inlet ports on the side panel Note Solvent Level Sensing and B Solvent Prime options should be disabled when completing the system configuration settings in se...

Страница 14: ...e carboy The small tube should extend 5 0 5 cm long from nut If the tube is missing insert it now and finger tighten the nut Figure 2 2 Optional Waste Cap 83B cap shown 4 The tubing is secured using m...

Страница 15: ...view the options below and complete as necessary If any optional modules are not being installed you can skip this section and continue to Section 2 13 2 12 1 Optional Sample Load Pump 1 Position the...

Страница 16: ...tion Guide Figure 2 4 Foxy R2 high flow diverter valve ports front view Use the supplied cable ties to secure the tubing as desired Before turning on any fraction collector manually move each fraction...

Страница 17: ...e Fractionation Valve was shipped with pairs of numbered clips to simplify tube and container identification If desired attach the clips to the each end of the six outlet tubes 5 Secure the six outlet...

Страница 18: ...Foxy R2 fraction collector Section 2 12 2 or the fractionation valve Section 2 6 3 Connect the ELSD to the DETECTOR port on the rear panel of the Foxy R2 or the fractionation valve with signal cable 2...

Страница 19: ...25 OD x 0 188 ID x 100cm long black PTFE tubing w nut and ferrule swaged one end and a metal union on the other end To plumb the ELSD to the Torrent 1 With a 3 32 Allen wrench remove the PEEK tee fitt...

Страница 20: ...bing Attach one end to the bottom of the metal tee fitting and the other end to the ELSD liquid inlet refer to Figure 2 11 Tighten fittings to finger tight 4 On the Torrent disconnect the black PTFE t...

Страница 21: ...cation all components along the fluid path must be electrically grounded An electrical ground is necessary to dissipate electrostatic charges that may build while collecting fluids such as organic sol...

Страница 22: ...to each unit Contact your IT department for assistance Note This section is dedicated to network configurations including the Foxy R2 fraction collector For comprehensive Torrent networking details re...

Страница 23: ...4 Enter Fraction Collector IP information into Torrent up to four units Torrent Foxy R2 Network Configuration Examples Typical examples of Torrent Foxy R2 network connection are provided in this secti...

Страница 24: ...ch unit connected to an Ethernet hub and then to your network The pass through function of a true Ethernet hub is required for this setup meaning all ports are sent the same signals Some switches can...

Страница 25: ...nlet line is receiving hexane and the Solvent B inlet is receiving ethyl acetate 4 Select the Tools Manual Control menu command to open the Manual Control window 5 Select the THROUGH COLUMN CARTRIDGE...

Страница 26: ...g the optional Solvent Selection Valve choose methanol or another suitable organic solvent as the B1 solvent and an additional organic solvent as the B2 solvent 4 Select the Tools Manual Control menu...

Страница 27: ...is recommended that the system operation be verified To do so follow the directions supplied with the Torrent Verification Kit Be sure to follow the correct instructions depending if the system is set...

Страница 28: ...SED ENSURE THE WASTE LINES ARE SECURED SO THEY DRAIN INTO THE WASTE CONTAINER SECURE THE WASTE LEVEL SENSE AIR TUBING SO ITS OUTLET IS AT LEAST TWO INCHES 5 CM BELOW THE CONTAINER S MAXIMUM LEVEL SEAL...

Страница 29: ...y Use Standard Description Severity Applied Performance Criteria EN61000 4 2 Electrostatic Discharge Level 2 4kV contact discharge Level 3 8kV air discharge A EN61000 4 3 Radiated RF Immunity 80 MHz t...

Страница 30: ...ct O ST O Represent the concentration of the hazardous substance in this component s any homogeneous pieces is lower than the ST standard limitation X ST X X Represent the concentration of the hazardo...