CombiFlash Torrent Installation Guide
1.3 Specifications
1.4 Safety
Before installing, operating, or maintaining this equipment, it is imperative that
all hazards and preventive measures are fully understood. While specific hazards
may vary according to location and application, take heed in the following general
warnings:
Table 1-1 Combi
Flash
Torrent Module Specifications
1
Sample size:
0.5 to 300 grams
Redi
Sep
Rf column
sizes:
80 gram to 3 kilogram
Flow rate:
50 – 1000 mL/min
Maximum pressure
2
:
6.89 bar (100 psi)
Detection:
200 – 400 nm (200 – 800 nm optional) with single,
dual, or All-wavelength Collection.
Accepts external detector input.
Gradient:
Binary, high pressure solvent mixing with linear, step,
or isocratic capability. Two solvent inlets.
Gradient accuracy:
<2% full scale from 0.2 to 1.0 liter/min.
Sample injection:
Automated injection with options of solid load
cartridges, liquid load, or direct injection.
Controller:
26.7 cm (10.4 inch) touch screen
Supports remote control via Windows® PC.
Software:
On-screen method development and control with
PeakTrak software
Fraction collection:
Up to four fraction collectors, fractionation valve, or
manual collection
Solvent management:
Patented, active level sensing for inlet and waste
containers
Module dimensions:
(W x D x H)
47 x 43.5 x 71.4 cm
18.5 x 17.9 x 28.1 inches
Module weight:
65 lbs (29.5 kg)
Operating voltage:
100 VAC 50 Hz, 2 amperes
117 VAC 60 Hz, 2 amperes
234 VAC 50 Hz, 1 ampere
(±10% of the region’s nominal line voltage)
Safety and sample
security:
Static-dissipative tubing throughout,
Vapor sensor for internal leak detection,
Overpressure sensor,
Active solvent and waste level sensing,
Audible alarm when user intervention is required.
Ambient Temperature
20 to 40°C (maximum temperature must be at least
15°C below the boiling point of the solvent)
Humidity (when
connected to power)
95% relative humidity maximum at 20 to 40°C
Certification:
CE
Electrical Safety per EN 61010-1
Installation Category
II
Maximum Altitude
2000 meters
Pollution Degree
2
Note 1. All specifications are subject to change.
Note 2. For columns undetected by RFID, the max pressure is limited to 50 psi.