CombiFlash Torrent Installation Guide
Figure 2-6 Six-port valve connections
2.13 Model 340CF Evaporative Light Scattering Detector (ELSD)
The Torrent can be configured for use with Teledyne Isco’s 340CF ELSD.
If a 340CF ELSD is no being installed you can skip this section and continue to
Section 2.14. The data connection between the Torrent and the ELSD requires either a
Foxy R2 fraction collector (Section 2.12.2) or the fractionation valve (Section 2.6.3).
Connect the ELSD to the DETECTOR port on the rear panel of the Foxy R2 or the
fractionation valve with signal cable (250000020) from the 340CF Accessory Package
(4-200203).
Note
If your system has more than one fraction collector, connect the ELSD to the first unit plumbed to
the Torrent.
Figure 2-7 Detector port on fractionation valve (left) and Foxy R2 fraction collector (right)
1
2
3
4
5
6
Collection Vessels
From
Torrent
Normally
Closed
Common
Normally
Open
Waste
From top of diverter valve