a. Locate a nut (60-0923-015) and ferrule (60-0083-163) in the accessory package
(60-5249-002).
b. Insert the ferrule, narrow end first, into the A LEVEL port.
c. Loosely thread the nut onto the A LEVEL port.
d. Insert the Solvent A air line into the A LEVEL port until it is fully seated.
e. Finger-tighten the nut. If desired, tighten the nut an additional
1
2
turn using a
wrench.
Important
For systems without Solvent Select Valve follow steps 4-7. For systems with Solvent
Select Valve skip to steps 8-14.
4. Locate the Solvent B inlet line. This assembly consists of three tubes fastened
together—a solvent line, a smaller diameter air line, and a solvent prime line.
5. Connect the large fitting on the Solvent B inlet line to the Torrent’s B INLET port as
described in step 2.
6. Connect the smallest tube of the Solvent B inlet line to the B LEVEL port as described
in step 3.
7. Connect the remaining tube of the Solvent B inlet line to the B PRIME port:
a. Finger-tighten the pre-swaged fitting to the port
labeled B PRIME.
b. Use a wrench to tighten the fitting an additional
1
8
turn.
8. The Solvent Select Valve should be plumbed to the B
Inlet of the Torrent (Section 2.2.6).
9. Locate the Sample Inlet line. This assembly consists of
two tubes fastened together--a white solvent line and a
smaller diameter air line.
10. Connect the large fitting on the Solvent B1 inlet line to
the Solvent Select Valve’s B1 INLET port as described in
step 2 above.
11. Connect the smalled tube of the Solvent B1 inlet line to
the Solvent Select Valve’s B Level port as described in
step 3.
12. Locate the Solvent B1 inlet line. This assembly consists
of two tubes fastened together--a solvent line and a
smaller diameter air line.
13. Connect the large fitting on the Solvent B1 inlet line to
the Solvent Select Valve’s B1 INLET port as described in
step 2 above.
14. Connect the smalled tube of the Solvent B1 inlet line to
the Solvent Select Valve’s B1 inlet line to the Solvent
Select Valve’s B Level port as described in step 3.
15. Locate the Solvent B2 inlet line and repeat Steps 9 and
10 for the B2 inlet line.
The other ends of the Solvent A and B lines should be
placed in the respective solvent containers. If using the
optional Solvent Inlet Cap sets (60-5394-510 for 83B
closures, 60-5394-511 for 100 mm closures), proceed with
the following steps:
16. Screw the cap onto the carboy.
Содержание CombiFlash Torrent