CombiFlash Torrent Installation Guide
Plumbing Setup:
• Connect the 107 cm (42 inch) piece of supplied tubing
between the Torrent’s TO COLLECTOR port and the
fraction collector’s diverter valve IN port (Common). Use
caution when tightening the hardware to prevent damage
to the diverter valve.
• If there are multiple fraction collectors, connect the 107 cm
(42 inch) piece of supplied tubing between the first fraction
collector’s diverter valve WASTE port (NO) and the next
collector’s diverter valve IN port (Common). Repeat this for
each fraction collector in the chain.
• On the last fraction collector (or single), connect the 6 ft.
piece of supplied tubing between the diverter valve WASTE
port (NO) and a suitable waste collection container. Cut the
tubing to the desired length.
• For waste level monitoring, the valve wash waste container
(section 2.10) is recommended for collecting diverter valve
waste.
• Install racks or funnels as described in the Installation
Guide.
Figure 2-4 Foxy R2 high flow diverter valve ports (front view)
• Use the supplied cable ties to secure the tubing as desired. Before turning on any
fraction collector, manually move each fraction collector’s arm and drop former
through its full range of movement to ensure that the arm mechanisms move freely
without obstruction. The Foxy R2 will need to be configured once the system is
powered on (Section 2.15 or 2.16).
2.12.3 Optional Fractionation Valve
The fractionation valve is an alternative to the typical Foxy
fraction collector setup, allowing the effluent from the Torrent
to be collected in any of six user-supplied containers. The
fractionation valve module includes a diverter valve to allow
the effluent to be diverted to waste.
1. Connect the supplied tubing between the Torrent’s TO
COLLECTOR port and the Fractionation Valve’s diverter
valve IN port (Common). (The diverter valve is an
externally mounted solenoid valve on the side of the
Fractionation Valve.)
2. Connect the supplied tubing between the Fractionation
Valve Common Port to the Fractionation Valve’s diverter
valve top port.
Waste (NO)
In (Common)
Drop Former (NC)