![Teledyne CombiFlash Torrent Скачать руководство пользователя страница 14](http://html1.mh-extra.com/html/teledyne/combiflash-torrent/combiflash-torrent_installation-manual_1080515014.webp)
CombiFlash Torrent Installation Guide
1. Select the tube with the pre-swaged nut and ferrule and connect this end to the port
labeled VALVE. The connection should be finger-tight. Then, use a wrench to tighten
the fitting an additional
1
8
turn.
2. Connect the other line to the LEVEL port:
a. Locate a nut (60-0923-015) and ferrule (60-0083-163) in
the accessory package (60-5249-002).
b. Insert the ferrule, narrow end first, into the LEVEL port.
c. Loosely thread the nut onto the LEVEL port.
d. Insert the waste air line into the LEVEL port until it is
fully seated.
e. Finger-tighten the nut. If desired, tighten the nut an
additional
1
2
turn using a wrench.
The other ends of the Waste line should be placed in the waste
container.
If using the optional Waste Outlet Cap (60-5394-512 for 83B closures, 60-5394-513 for 100
mm closures), proceed with the following steps:
3. Place the metal plate inside the plastic cap (Figure 2-2). The plate should be oriented so
that the small tube will extend into the carboy. The small tube should extend 5 ±0.5 cm
long from nut. If the tube is missing, insert it now and finger-tighten the nut.
Figure 2-2 Optional Waste Cap (83B cap shown)
4. The tubing is secured using metal, spring-type retainers. A pair of conventional pliers
will allow you to open the retainer enough to slide it onto the tubing. On the larger
diameter waste tube, slide one retainer about 2 inches (5 cm) onto the tubing end.
5. Feed this tube into the top of the metal plate until the retainer is against the top of the
plate.
6. Slide a retainer onto the tubing end until it is seated against the bottom of the metal
plate.
7. Thread the cap onto the carboy.
8. The smaller diameter tube is the level sense tube. Push this tube about 2 cm into the
top fitting and finger-tighten the nut.
If an optional waste cap is not used, ensure the waste lines are secured so they drain into
the waste container. Secure the Waste Level Sense air tubing so its outlet is at least two
inches (5 cm) below the container’s maximum level. Seal the container opening to avoid
solvent vapors.