Section 2 Preparation
Figure 2-17 Network configurations: Example 3
2.17 Purge the Internal Solvent Lines (Torrent)
If setting up the Torrent AQ skip to section 2. The Torrent is shipped from the factory
with isopropyl alcohol in the internal lines. Before the first use, the lines should be
purged. To do so:
1. Connect the solid load cartridge tubing to the sample injection port (Figure 1-1).
CAUTION
Pinch point. Your fingers or hands can be seriously injured if you place them
between the moving parts of the Solid Load Cartridge Cap Mount or the Upper
Column Mount. Use care when installing solid load cartridges or columns.
2. Insert the column bypass tube (part number 209-0165-76 from accessory kit
60-5249-002).
3. Ensure that the Solvent A inlet line is receiving hexane, and the Solvent B inlet is
receiving ethyl acetate.
4. Select the “Tools>Manual Control” menu command to open the Manual Control
window.
5. Select the T
HROUGH
C
OLUMN
& C
ARTRIDGE
Valve Position setting.
6. Touch the P
RIME
B button. The solvent pump will start.
Note
When pumping solvents the first time, inspect the solvent and waste connections to the
system. If any leaks are observed, click the S
TOP
button. Correct the leak by tightening the
fitting an additional
1
4
turn, and then restart the pump by touching the P
RIME
B button.
7. After at least 400 mL of solvent has been pumped, touch the S
TOP
button.
8. Touch the P
RIME
A button. The solvent pump will start. Inspect the solvent and
waste lines and correct leaks if necessary.
9. After at least 400 mL of solvent has been pumped, touch the S
TOP
button.
10. Close the Manual Control window.
Your network
Ethernet hub
Foxy R2
rear panel
Torrent rear panel
Foxy R2
rear panel