Section 2 Preparation
2.11 Connect and Route Drain Line
WARNING
Risk of fire or equipment damage. Failure to connect drain lines may allow organic
solvents to pool in unsafe areas, creating a potential for dangerous levels of flammable
vapors. Improper draining may damage the instrument’s internal components.
The Torrent has a front-mounted drip tray. There is a drain connection under the left
side of the tray. Connect user-supplied tubing to this drain and route it to a suitable
waste fluid collection container.
2.12 Plumbing Optional Modules to Torrent
The installation of optional equipment that was electrically connected to the Torrent in
section 2.6 should be completed now. Review the options below and complete as
necessary. If any optional modules are not being installed you can skip this section and
continue to Section 2.13.
2.12.1 Optional Sample Load Pump
1. Position the Sample Load Pump on top or near the
Torrent.
2. Connect the tubing between the pump outlet (from
the bottom of the pump head) and the Torrent’s
sample injection port (Figure 1-1, item 9).
3. Configure the Sample Load Pump by placing either
the plastic cartridge reservoir or the tubing with the
weighted end on the pump inlet (Figure 2-3).
Figure 2-3 Sample load pump
2.12.2 Optional Foxy R2 High Flow Fraction Collector
The fraction collector(s) should be positioned to the right of the Torrent. Complete
information about the Foxy R2 fraction collector is provided in its user manual, part
#69-2133-667.
Pump Outlet
To Sample Injection Port
Cartridge Reservoir
Tubing with
Weighted End
Pump Inlet
Содержание CombiFlash Torrent
Страница 6: ...CombiFlash Torrent Installation Guide...