Installation Guide
Section 2 Preparation
This section provides instructions for unpacking and installing the Combi
Flash
Torrent
system. To prepare the system for operation, sequentially follow all instructions in
sections 2.1 through 2.18. Some of these sections provide instructions for installing
optional equipment which should only be completed as required.
Note
Section
contains an Installation Qualification checklist. If required, sign off the
checklist entries as you successfully complete the following sections.
2.1 Optional Equipment Modules
The Torrent/Torrent AQ offer several different optional modules. Each module will
require additional set-up steps as noted below. The optional modules are:
1. Solvent Selection Valve (Sections 2.4.1; 2.6.2 and 2.8)
2. Sample Load Pump (Sections 2.6.1 and 2.12.1)
3. Fractionation Valve (Sections 2.4.2; 2.6.3 and 2.12.3)
4. Foxy R2 High Flow Fraction Collector (Sections 2.6.4; 2.12.2 and/or 2.17)
2.2 Unpacking the Unit
The Torrent is shipped in a single carton. The optional equipment will be shipped in
additional cartons. Carefully unpack the shipment and inspect the contents.
WARNING
The Torrent is heavy. Use a two-person lift to prevent injury.
Lift the Torrent from the container using the lifting handles mounted on the
side panels.
If there is damage to the shipping carton or any components, contact the
shipping agent and Teledyne Isco (or its authorized representative)
immediately.
WARNING
If there is any evidence that the Torrent has been damaged in shipping, do not plug it
into AC power. Contact Teledyne Isco or its authorized representative for advice.
Compare the contents of the boxes with the enclosed packing slips. If there are any
shortages, contact Teledyne Isco immediately.
2.3 Instrument Location
The Torrent has a relatively small footprint for a development-scale instrument,
requiring about 2200 square centimeters (350 in
2
) of level bench space. Ensure that the
Torrent has at least 3 cm (1.25") of air space behind it for ventilation. Additional space
may be required for solvent and waste containers, and fraction collection.
Refer to Table 1-1 for environmental conditions and power requirements.