
ADFM Test Procedures
Chapter
5
ADFM Test Procedures
5-1 Introduction
This chapter explains how to test the ADFM using
BBTALK
and the
WinADFM
program. These tests thoroughly check the ADFM in a
laboratory environment, but are no substitute for a practice deployment.
You should test the ADFM:
•
When you first receive the ADFM.
•
Before each deployment or every six months.
•
When you suspect instrument problems.
•
After each deployment.
These test procedures assume all equipment is working. The tests can help
you isolate problems to a major functional area of the ADFM. For trouble-
shooting information, see
Chapter 6
.
NOTE.
The built-in tests require you to immerse the transducer faces in
water. If you do not, some of the tests may fail. Running the tests in air will
not harm the ADFM.
5-2 Test
Setup
Use the following steps to connect the ADFM system and to place the
ADFM in a known state.
a. Connect a laptop computer to the RS-232 communication port.
b. Connect and apply power to the system as described in
Chapter 2
.
c. Place the ADFM transducer in approximately one foot of water.
ADFM Technical Manual (January 2000)
page 5-1
Содержание ADFM Analog Output Module
Страница 2: ......
Страница 6: ...Table of Contents NOTES iv MGD Technologies Inc ...
Страница 13: ...Table of Contents NOTES ADFM Technical Manual January 2000 xi ...
Страница 14: ...Table of Contents xii MGD Technologies Inc ...
Страница 54: ...Chapter 5 NOTES page 5 10 MGD Technologies Inc ...
Страница 132: ...Appendix C NOTES page C 52 MGD Technologies Inc ...
Страница 152: ...Appendix E NOTES page E 14 MGD Technologies Inc ...
Страница 154: ......
Страница 155: ......
Страница 156: ......