ADFM Commands
deep tracking. This allows the ADFM to change with condi-
tions during the course of the deployment.
VN - Vertical Beam Number Of Depth Cells
Purpose:
Sets the number of depth cells over which the ADFM collects
data for the vertical beam.
Format: VN
nnn
Range:
nnn
= 001 to 128 depth cells
Description: The range of the ADFM is set by the number of depth cells
(VN) times the size of each depth cell (VS).
VP - Pings per Ensemble
Purpose:
Sets the number of pings for the vertical beam to average in
each data ensemble.
Format: VP
nnnnn
Range:
nnnnn
= 0 to 16384 pings
Default: VP00010
Description: VP sets the number of vertical beam pings to average in each
ensemble before sending/recording the water-track data.
VS - Vertical Beam Depth Cell Size
Purpose:
Selects the volume of water for one measurement cell for the
vertical beam
Format: VS
nnnn
Range:
nnnn
= 0 to 9999 cm
Description: The ADFM collects data over a variable number of depth
cells. VS sets the size of each cell in vertical centimeters
VX - Vertical Beam Transmit Length
Purpose:
Sets the vertical beam transmit length in centimeters.
Format: VX
nn
Range:
nn
= 0 to 99 cm
Description: Sets the vertical beam transmit length in centimeters.
ADFM Technical Manual (January 2000)
page C-47
Содержание ADFM Analog Output Module
Страница 2: ......
Страница 6: ...Table of Contents NOTES iv MGD Technologies Inc ...
Страница 13: ...Table of Contents NOTES ADFM Technical Manual January 2000 xi ...
Страница 14: ...Table of Contents xii MGD Technologies Inc ...
Страница 54: ...Chapter 5 NOTES page 5 10 MGD Technologies Inc ...
Страница 132: ...Appendix C NOTES page C 52 MGD Technologies Inc ...
Страница 152: ...Appendix E NOTES page E 14 MGD Technologies Inc ...
Страница 154: ......
Страница 155: ......
Страница 156: ......