Appendix C
WN - WT Number Of Depth Cells
Purpose:
Sets the number of depth cells over which the ADFM collects
data.
Format: WN
nnn
Range:
nnn
= 001 to 128 depth cells
Description: The range of the ADFM is set by the number of depth cells
(WN) times the size of each depth cell (WS).
WO - Single Beam Mode Select
Purpose:
Selects a single beam to ping from.
Format: WO
n
Range:
n
= 0 to 5 (0 = off, 1-5)
Default: WO0
Description: Selects a single beam to ping from. If the single beam com-
mand is not zero all water profiling and vertical pings come
from the same ceramic transducer.
WP - WT Pings Per Ensemble
Purpose:
Sets the number of pings to average in each data ensemble.
Format: WP
nnnnn
Range:
nnnnn
= 0 to 16384 pings
Description: WP sets the number of water-track pings to average in each
ensemble before sending/recording the water-track data.
WS - WT Depth Cell Size
Purpose:
Selects the volume of water for one measurement cell.
Format: WS
nnnn
Range:
nnnn
= 0 to 9999 cm (328 feet) measured radially from each
beam.
Description: The ADFM collects data over a variable number of depth
cells. WS sets the size of each cell in vertical centimeters.
page C-50
MGD Technologies, Inc.
Содержание ADFM Analog Output Module
Страница 2: ......
Страница 6: ...Table of Contents NOTES iv MGD Technologies Inc ...
Страница 13: ...Table of Contents NOTES ADFM Technical Manual January 2000 xi ...
Страница 14: ...Table of Contents xii MGD Technologies Inc ...
Страница 54: ...Chapter 5 NOTES page 5 10 MGD Technologies Inc ...
Страница 132: ...Appendix C NOTES page C 52 MGD Technologies Inc ...
Страница 152: ...Appendix E NOTES page E 14 MGD Technologies Inc ...
Страница 154: ......
Страница 155: ......
Страница 156: ......