ADFM Commands
WV - WT Mode 1 Ambiguity Velocity
Purpose:
Sets the radial ambiguity velocity for profiling Mode 1
(WM1) and the Mode 1 portion of profiling Mode 4 (WM4).
Format: WV
nnn
Range:
nnn
= 020 to 999 cm/s
Description: Set WV as low as possible to attain maximum performance,
but not too low or ambiguity errors will occur. Rule of thumb: Set WV to
the maximum relative horizontal velocity between water-current speed and
ADFM speed.
ADFM Technical Manual (January 2000)
page C-51