ADFM Commands
255 255
0 0
0 0
255 255
0 0
0 0
0 0
255 255
Electronics Test Results = $00000000 ... PASS
PT5 Failure Description
You can determine failure results ($>0) by the individual bit settings:
BIT #
PT5 Failure Description
0
Beam 1 Failure - A high value (normally 255) was <254, or a low value
(normally 0) was >20.
1
Beam 2 Failure - See Bit 0
2
Beam 3 Failure - See Bit 0.
3
Beam 4 Failure - See Bit 0.
4
Correlator Timeout - The CPU never received a "processing done" signal
from the Correlator.
PT6 – Receive Bandwidth
Purpose:
This test measure the receive bandwidth of the system.
Format: PT6
Description: This test measure the receive bandwidth of the system. The
bandwidth varies with system frequency and the WB com-
mand setting. Beam fails if <75% or >125% of nominal value.
Example:
>pt6
Receive Bandwidth: Narrow
Sample bw
rate expect Bm1 Bm2 Bm3 Bm4 Bm5
307200 45 31 36 34 35 26 Khz
results PASS PASS PASS PASS PASS
Receive Bandwidth: Middle
Sample bw
rate expect Bm1 Bm2 Bm3 Bm4 Bm5
614400 89 60 73 68 71 52 Khz
results PASS PASS PASS PASS PASS
Receive Bandwidth: Wide
Sample bw
rate expect Bm1 Bm2 Bm3 Bm4 Bm5
1228800 178 125 145 138 143 102 Khz
results PASS PASS PASS PASS PASS
ADFM Technical Manual (January 2000)
page C-31
Содержание ADFM Analog Output Module
Страница 2: ......
Страница 6: ...Table of Contents NOTES iv MGD Technologies Inc ...
Страница 13: ...Table of Contents NOTES ADFM Technical Manual January 2000 xi ...
Страница 14: ...Table of Contents xii MGD Technologies Inc ...
Страница 54: ...Chapter 5 NOTES page 5 10 MGD Technologies Inc ...
Страница 132: ...Appendix C NOTES page C 52 MGD Technologies Inc ...
Страница 152: ...Appendix E NOTES page E 14 MGD Technologies Inc ...
Страница 154: ......
Страница 155: ......
Страница 156: ......