3700 Portable Samplers
Section 3 Programming Guidelines
3-27
Example
Flow-Paced Sequential Sampling
shows how to
program the sampler to take flow-paced samples at intervals of 5
pulses. One 500 ml sample is to be placed in each bottle. The
sampling program will default to the value entered in Start Time
Delay option.
8.
Because this sampling routine does not require a specific start time, use the arrow keys to select “NO.” Press the
ENTER
/
PROGRAM
key to accept the entry. If you select “YES,” you can enter a specific start time and date for the
routine.
ENTER START TIME?
[YES,
NO
]
After this message displays briefly, the sampler will automatically return to the standby state.
PROGRAMMING SEQUENCE
COMPLETE
9.
After you properly install the sampler, press the
START
SAMPLING
key to run the program.
. . . STANDBY . . .
5:42:23 19-JUN-04
10. To start the sampling routine with the first bottle, accept the blinking “1” by pressing the
ENTER
/
PROGRAM
key. If
you want to start the routine with another bottle, enter the starting bottle number.
START SAMPLING
BOTTLE
1
(1-24)
This display appears as the sampler counts down the time remaining to the start time. The first line reports the bottle
which will receive the next sample volume. The second line reports the start time at the left and the current time on
the right.
BOTTLE 1
AT 5:44
5:42:33
When the start time arrives, the sampler will take the first sample. The sample event cycle begins with a pre-sample
purge. During the purge, the display indicates the number of the bottle that will receive the sample volume.
BOTTLE 1
As soon as the pump runs forward to deliver the sample volume, the message on the second line appears. This
message remains through the post-sample purge.
BOTTLE 1
PUMPING 250 ml
This display appears when the post-sample purge from the previous sample event is completed. It indicates the
number of the bottle that is to receive the sample at the upcoming sample event. The left side of the second line
displays the scheduled time for the next event. The current time is displayed on the right side.
BOTTLE 2
AT 6:14
5:45:11
The cycle of displays, shown in Steps 12 through 13, is repeated for each bottle until the sampling routine is done.
↓
↓
When the routine is completed, this message appears. It reports the status of the routine (“DONE”), the total number
of sample events, and the current time and date.
DONE . . . 24 SAMPLES
6:10:35 21-JUN-04
Example: Time-Paced Sequential Sampling (Continued)
Содержание 3700
Страница 2: ......
Страница 4: ......
Страница 33: ...3700 Portable Samplers Section 2 Setup Procedures 2 9 Figure 2 7 Locked Sampler ...
Страница 34: ...3700 Portable Samplers Section 2 Setup Procedures 2 10 ...
Страница 99: ...3700 Portable Samplers Section 3 Programming Guidelines 3 65 Figure 3 14 Start Time Diagram I I I I I I ...
Страница 144: ...3700 Portable Samplers Section 3 Programming Guidelines 3 110 ...
Страница 152: ...3700 Portable Samplers Section 4 Options and Interfacing Equipment 4 8 ...
Страница 164: ...3700 Portable Samplers Section 5 Routine Maintenance 5 12 ...
Страница 175: ...3700 Portable Samplers Section 6 Servicing 6 11 Figure 6 6 Main Circuit Board Component Layout OUT OUT OUT OUT OUT OUT ...
Страница 180: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 2 ...
Страница 181: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 3 ...
Страница 182: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 4 ...
Страница 183: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 5 ...
Страница 184: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 6 60 3703 262 REV J 3 2 ...
Страница 186: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 8 ...
Страница 188: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 10 ...
Страница 189: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 11 ...
Страница 190: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 12 ...
Страница 191: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 13 ...
Страница 192: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 14 ...
Страница 193: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 15 ...
Страница 194: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 16 ...
Страница 195: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 17 ...
Страница 196: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 18 ...
Страница 200: ...3700 Portable Samplers Appendix B Accessories B 4 ...
Страница 210: ...3700 Portable Samplers Appendix C General Safety Procedures C 10 ...
Страница 224: ...3700 Portable Samplers Appendix E Calculating Flow Increment Between Samples E 4 ...
Страница 234: ...3700 Portable Samplers Appendix G Material Safety Data Sheets G 8 ...
Страница 238: ......
Страница 240: ......
Страница 242: ......