3700 Portable Samplers
Section 3 Programming Guidelines
3-25
15. Select “ENABLE” to turn on the Liquid Detector on, “DISABLE” to turn off the Liquid Detector. For the purposes of
the following examples, select “ENABLE.” Press the
ENTER
/
PROGRAM
key to accept the selection. You should
normally leave the detector enabled unless you suspect it is malfunctioning. If you disable the detector, you will
need to enter the suction head in the program sequence.
[
ENABLE
, DISABLE]
LIQUID DETECTOR
16. This display appears when you select “ENABLE” in step
15
. For the purposes of this example, enter “0.” Press the
ENTER
/
PROGRAM
key to accept the entry. Rinse cycles condition the suction line to reduce cross contamination.
0
RINSE CYCLES (0-3)
17. This display appears when you select “ENABLE” in step
15
. For the purposes of the following examples, select
“NO” to omit the setting. Press the
ENTER
/
PROGRAM
key to accept the selection.
ENTER HEAD MANUALLY?
[YES,
NO
]
18. This display appears when you select “ENABLE” in step 15. For the purposes of this example, set the number of
retries to “0.” Press the
ENTER
/
PROGRAM
key to accept the entry. This setting determines the number of times the
sampler will try to detect the presence of liquid for each sample event.
RETRY UP TO
0
TIMES
WHEN SAMPLING (0-3)
19. To verify the programming mode setting, press the
ENTER
/
PROGRAM
key.
SELECT OPTION: (
← →
)
PROGRAMMING MODE
20. Select “BASIC.” Press the
ENTER
/
PROGRAM
key to accept the selection
[
BASIC
, EXTENDED]
PROGRAMMING MODE
21. Scroll through the remaining options with the arrow keys. Use the techniques above to change or verify the
settings. If you want to reinitialize the configure and program settings to factory settings, follow the procedures
given in Steps
22 to 24.
↓
↓
22. To access the displays used in this option, press the
ENTER
/
PROGRAM
key.
SELECT OPTION: (
← →
)
RUN DIAGNOSTICS
The sampler displays the software revision number for a short period of time.
SOFTWARE REVISION
#4.6
The software revision display will be replaced by the RAM and ROM test messages. A successful test is indicated by
the messages “RAM PASSED TEST” and “ROM PASSED TEST.”
‘RAM’ PASSED TEST
TESTING ‘ROM’
After successful RAM and ROM tests, the sampler will then test the LCD by first filling the display with solid rectangles
and then printing the alphabet and other characters. Each position in the display should contain a character.
ABCDEFGHIJKLMNOPQRST
UVWXYZ [¥]^_`abcdefgh
The next step tests the pump. During the test, the pump will run briefly and the display will indicate an “OFF/ON”
number. The number should fall within the range of 50 to 200. A count near 100 is typical.
Example: Checking the Configure Option Settings (Continued)
Содержание 3700
Страница 2: ......
Страница 4: ......
Страница 33: ...3700 Portable Samplers Section 2 Setup Procedures 2 9 Figure 2 7 Locked Sampler ...
Страница 34: ...3700 Portable Samplers Section 2 Setup Procedures 2 10 ...
Страница 99: ...3700 Portable Samplers Section 3 Programming Guidelines 3 65 Figure 3 14 Start Time Diagram I I I I I I ...
Страница 144: ...3700 Portable Samplers Section 3 Programming Guidelines 3 110 ...
Страница 152: ...3700 Portable Samplers Section 4 Options and Interfacing Equipment 4 8 ...
Страница 164: ...3700 Portable Samplers Section 5 Routine Maintenance 5 12 ...
Страница 175: ...3700 Portable Samplers Section 6 Servicing 6 11 Figure 6 6 Main Circuit Board Component Layout OUT OUT OUT OUT OUT OUT ...
Страница 180: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 2 ...
Страница 181: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 3 ...
Страница 182: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 4 ...
Страница 183: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 5 ...
Страница 184: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 6 60 3703 262 REV J 3 2 ...
Страница 186: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 8 ...
Страница 188: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 10 ...
Страница 189: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 11 ...
Страница 190: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 12 ...
Страница 191: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 13 ...
Страница 192: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 14 ...
Страница 193: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 15 ...
Страница 194: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 16 ...
Страница 195: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 17 ...
Страница 196: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 18 ...
Страница 200: ...3700 Portable Samplers Appendix B Accessories B 4 ...
Страница 210: ...3700 Portable Samplers Appendix C General Safety Procedures C 10 ...
Страница 224: ...3700 Portable Samplers Appendix E Calculating Flow Increment Between Samples E 4 ...
Страница 234: ...3700 Portable Samplers Appendix G Material Safety Data Sheets G 8 ...
Страница 238: ......
Страница 240: ......
Страница 242: ......