3700 Portable Samplers
Section 3 Programming Guidelines
3-98
Manual Sample –
You can take manual samples in standby by
pressing the
MANUAL
SAMPLE
key. After you press the
MANUAL
SAMPLE
key, the sampler will deliver the programmed sample
volume to the bottle currently under the distributor. A manual
sample delivers the sample volume currently entered as a pro-
grammed setting, and includes the pre-sample and post-sample
purges. It will also include any programmed rinses and retries.
The sampling process can be stopped with the
STOP
key at any
time.
While the sample is being delivered, the sampler displays a
message similar to that illustrated below. The second line will
not appear until the pump begins to pump forward to deliver the
sample. The display will revert to standby after the manual
sample has been delivered.
Next Bottle
- Pressing the
NEXT
BOTTLE
key will cause the dis-
tributor to move to the next bottle. If the distributor is positioned
over the last bottle, it will move to bottle position 1. The sampler
will display the distributor’s location as illustrated below. After a
short time, the display will revert to standby.
Start Sampling –
Once you have programmed and configured
the sampler, use the
START
SAMPLING
key to run the sampling
program. The first display you’ll see after pressing the
START
SAMPLING
key is illustrated below.
If you want to have the sampling program begin with a bottle
other than Bottle 1, enter the first bottle number here. If “1” is
acceptable, press the
ENTER
/
PROGRAM
key to place the sampler
into the run state. If you do not press a key within 60 seconds,
the sampler will automatically enter the run state and will begin
the sampling routine with Bottle 1.
There are three instances where the sampler will not be able to
run the program immediately. These occur when the program
has been started after: 1) the programmed start time, 2) the first
switch time, and 3) one or more programmed stop times. The
sampler responds to these situations as discussed in Examples:
Program Started Later Than Programmed Start Time
Started Later Than Programmed 1st Switch Time
,
and
Started Later Than Programmed Stop Time
.
Resume Sampling –
To resume the program, press the
RESUME
SAMPLING
key. When you press
RESUME
SAMPLING
, the program
will resume at the point it was halted. The display must read
“PROGRAM HALTED” for the resume sampling key to be active.
If any samples were missed during the time the program was
. . . MANUAL SAMPLE . . .
PUMPING 200 ml
NOW ON BOTTLE
2
START SAMPLING
AT BOTTLE
1
(1 - 24)
Содержание 3700
Страница 2: ......
Страница 4: ......
Страница 33: ...3700 Portable Samplers Section 2 Setup Procedures 2 9 Figure 2 7 Locked Sampler ...
Страница 34: ...3700 Portable Samplers Section 2 Setup Procedures 2 10 ...
Страница 99: ...3700 Portable Samplers Section 3 Programming Guidelines 3 65 Figure 3 14 Start Time Diagram I I I I I I ...
Страница 144: ...3700 Portable Samplers Section 3 Programming Guidelines 3 110 ...
Страница 152: ...3700 Portable Samplers Section 4 Options and Interfacing Equipment 4 8 ...
Страница 164: ...3700 Portable Samplers Section 5 Routine Maintenance 5 12 ...
Страница 175: ...3700 Portable Samplers Section 6 Servicing 6 11 Figure 6 6 Main Circuit Board Component Layout OUT OUT OUT OUT OUT OUT ...
Страница 180: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 2 ...
Страница 181: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 3 ...
Страница 182: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 4 ...
Страница 183: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 5 ...
Страница 184: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 6 60 3703 262 REV J 3 2 ...
Страница 186: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 8 ...
Страница 188: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 10 ...
Страница 189: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 11 ...
Страница 190: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 12 ...
Страница 191: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 13 ...
Страница 192: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 14 ...
Страница 193: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 15 ...
Страница 194: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 16 ...
Страница 195: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 17 ...
Страница 196: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 18 ...
Страница 200: ...3700 Portable Samplers Appendix B Accessories B 4 ...
Страница 210: ...3700 Portable Samplers Appendix C General Safety Procedures C 10 ...
Страница 224: ...3700 Portable Samplers Appendix E Calculating Flow Increment Between Samples E 4 ...
Страница 234: ...3700 Portable Samplers Appendix G Material Safety Data Sheets G 8 ...
Страница 238: ......
Страница 240: ......
Страница 242: ......