3700 Portable Samplers
Section 3 Programming Guidelines
3-50
If the bottle sets were switched every two hours and the sampler
has been inhibited for six hours, the restarted routine will begin
with the next bottle set. The routine may not continue with the
bottle set that would have been used had the sampler not been
inhibited.
3.9.15 Event Mark
The Event Mark option configures the sampler to send an event
mark to an interfaced flow meter or other equipment. The
sampler will supply a variable duration pulse of up to 700 mA at
12 V on pin E of the flow meter connector. Four types of signals
can be sent:
• a pulse at the beginning of the pre-sample purge
• a pulse at the beginning of forward pumping only
• a continuous signal during the entire pump cycle
• a continuous signal during forward pumping only
The event mark includes a signal sent on pin D which is inter-
preted and recorded by an Isco flow meter as the bottle
number(s) for the sample event.
The bottle number signal consists of a number of 100 millisecond,
12-volt pulses that equal the bottle number;
i.e.
, four pulses rep-
resent bottle #4, 20 pulses represent bottle #20.
The type of signal selected from the Event Mark configure option
affects the event mark signal on pin E. In Figure 3-9A, a pulse
signal is sent at the beginning of the pre-sample purge. In Figure
3-9B, a pulse signal is sent during the sample volume delivery
portion of the sample event. In Figure 3-9C, a continuous signal
is sent during the entire pump cycle. In Figure 3-9D, a con-
tinuous pulse is sent for the duration of the sample-volume
delivery.
Содержание 3700
Страница 2: ......
Страница 4: ......
Страница 33: ...3700 Portable Samplers Section 2 Setup Procedures 2 9 Figure 2 7 Locked Sampler ...
Страница 34: ...3700 Portable Samplers Section 2 Setup Procedures 2 10 ...
Страница 99: ...3700 Portable Samplers Section 3 Programming Guidelines 3 65 Figure 3 14 Start Time Diagram I I I I I I ...
Страница 144: ...3700 Portable Samplers Section 3 Programming Guidelines 3 110 ...
Страница 152: ...3700 Portable Samplers Section 4 Options and Interfacing Equipment 4 8 ...
Страница 164: ...3700 Portable Samplers Section 5 Routine Maintenance 5 12 ...
Страница 175: ...3700 Portable Samplers Section 6 Servicing 6 11 Figure 6 6 Main Circuit Board Component Layout OUT OUT OUT OUT OUT OUT ...
Страница 180: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 2 ...
Страница 181: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 3 ...
Страница 182: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 4 ...
Страница 183: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 5 ...
Страница 184: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 6 60 3703 262 REV J 3 2 ...
Страница 186: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 8 ...
Страница 188: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 10 ...
Страница 189: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 11 ...
Страница 190: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 12 ...
Страница 191: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 13 ...
Страница 192: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 14 ...
Страница 193: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 15 ...
Страница 194: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 16 ...
Страница 195: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 17 ...
Страница 196: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 18 ...
Страница 200: ...3700 Portable Samplers Appendix B Accessories B 4 ...
Страница 210: ...3700 Portable Samplers Appendix C General Safety Procedures C 10 ...
Страница 224: ...3700 Portable Samplers Appendix E Calculating Flow Increment Between Samples E 4 ...
Страница 234: ...3700 Portable Samplers Appendix G Material Safety Data Sheets G 8 ...
Страница 238: ......
Страница 240: ......
Страница 242: ......