Testing A Fiber
TFS3031 TekRanger/TekRanger 2 User Manual
2–21
Real-Time and High-Density Testing
For real-time and high-density testing, the Fiber Scan setup in the
Test Setup menu must be set to Manual mode, and the waveform
should be zoomed.
Real-Time Testing. Real-time testing is useful when adjusting fiber
connections and making splices because it immediately displays the
results of work being done. New data is acquired quickly and the
waveform is continually updated. Changes appear quickly on the
screen.
Real-time testing can be done on either the full or zoomed
waveform. However, to see more detail and for faster updating of the
display, zoom the waveform.
CAUTION.
In real-time mode the laser fires continuously. Do not
operate the TFS3031 in real-time mode unless a fiber is connected to
the Laser Output port. Severe damage to internal electronics can
result.
High-Density Testing. High-density testing tests a specific section of
the fiber at higher density (the sample points are closer together).
Test Procedure
Step 1. Put the TFS3031 in Manual fiber test mode.
1. Push the
Setup
softkey.
2. Change the Fiber Scan setup in the Test Setup menu to
Manual.
Step 2. Push the
START/STOP
button to complete a test and
display the waveform.
Step 3. Position the active cursor on the event (or other area of the
waveform) that you want to test.
Testing A Fiber
TFS3031 TekRanger/TekRanger 2 User Manual
2–21
Real-Time and High-Density Testing
For real-time and high-density testing, the Fiber Scan setup in the
Test Setup menu must be set to Manual mode, and the waveform
should be zoomed.
Real-Time Testing. Real-time testing is useful when adjusting fiber
connections and making splices because it immediately displays the
results of work being done. New data is acquired quickly and the
waveform is continually updated. Changes appear quickly on the
screen.
Real-time testing can be done on either the full or zoomed
waveform. However, to see more detail and for faster updating of the
display, zoom the waveform.
CAUTION.
In real-time mode the laser fires continuously. Do not
operate the TFS3031 in real-time mode unless a fiber is connected to
the Laser Output port. Severe damage to internal electronics can
result.
High-Density Testing. High-density testing tests a specific section of
the fiber at higher density (the sample points are closer together).
Test Procedure
Step 1. Put the TFS3031 in Manual fiber test mode.
1. Push the
Setup
softkey.
2. Change the Fiber Scan setup in the Test Setup menu to
Manual.
Step 2. Push the
START/STOP
button to complete a test and
display the waveform.
Step 3. Position the active cursor on the event (or other area of the
waveform) that you want to test.
Содержание ATEC TekRanger 2
Страница 4: ......
Страница 18: ...Preface xiv TFS3031 TekRanger TekRanger 2 User Manual Preface xiv TFS3031 TekRanger TekRanger 2 User Manual ...
Страница 20: ......
Страница 68: ......
Страница 124: ......
Страница 132: ......
Страница 138: ...Cleaning 4 6 TFS3031 TekRanger TekRanger 2 User Manual Cleaning 4 6 TFS3031 TekRanger TekRanger 2 User Manual ...
Страница 146: ......
Страница 165: ...Index Index ...
Страница 166: ......
Страница 174: ...Index Index 8 TFS3031 TekRanger TekRanger 2 User Manual Index Index 8 TFS3031 TekRanger TekRanger 2 User Manual ...