Installation

Up to six NC415AV Master Stations may be connected to a single NC120 CE. Connect up to three mas-
ters to each 8P8C jack labeled master station on the front of the NC120 CE.

Multicolor Zone Dome

If desired, a zone dome light (part number LI484P5) may be connected to the NC415AV Master Station
using the 6P6C jack on the back. The zone dome light will illuminate whenever a call is present on the
connected master station.

Note

: Using a LI484P5 reduces the maximum number of NC415AVs on the module.

Use CAT5 or better cabling to connect the LI484P5 to the NC415AV Master Station and wire the 6P6C
with a straight-through configuration. Do not exceed 50 feet of cabling for dome light connection.

Connect Data Bus Wiring

Before connecting the data bus wiring to the NC120 CE Module, check the data bus wiring for shorts.
Connect the positive wire in the data bus cable to the positive terminal on the NC120 CE Module port
and the negative wire in the data bus cable to the negative terminal on the NC120 CE Module using the
included plugs.

Note which station run gets plugged into which CE module port. This information will be needed when
setting up the system configuration.

Connect Master Bus Wiring

Connect the master bus cabling to the ports on the right side of the NC120 Central Equipment Module.

Keep a record of the locations of master stations and their automatically assigned names. Once the sys-
tem is powered on for the first time, go to each master station on the system and touch the

Menu

button

on the home screen, then touch the

This Master

button that appears. The top of the screen will display

the default name of the master station. Note this information, as well as the actual location of the master
station for use during system configuration.

If more than one NC415AV Master Station is installed on a master bus run, use a CT400P5 splitter to
provide multiple connections to a single master bus port. Set the EOL resistor DIP switch on the last
NC415AV Master Station on the cable run to ON as shown in

Figure 4 on page 13

.

Networking Multiple NC120 CE Modules

If multiple CE modules are to be networked together, use the DIP switches on the face of the CE mod-
ules to set unique module IDs for each CE module as discussed in

The NC120 Central Equipment Mod-

ule on page 10

. Once each CE module has been assigned a unique address, connect the CE modules to

each other using the Ethernet jacks on the left side using CAT5 or better cabling terminated using 8P8C
connectors using the T568B standard.

Maximum cable length is 100 meters (328 feet). Extend this distance by placing NC554/8 network
switches every 100 meters. Use NC556 Fiber to Ethernet Converters to extend this distance up to 1,200
meters over a fiber optic connection.

Once the CE modules are networked together, proceed to the Configuration Setup section.

20

|

IL992 Tek-CARE120 Installation Manual

Copyright

©

TekTone Sound and Signal Mfg., Inc. All Rights Reserved

Содержание Tek-CARE120

Страница 1: ...Free 800 327 8466 Sales Option 2 Tech Support Option 3 324 Industrial Park Rd Franklin NC 28734 tektone tektone com Fax 828 524 9968 IL992 Rev 9 8 2020 Tek CARE120 Tone Visual Nurse Call Signaling Sy...

Страница 2: ...Tone TekTone Sound Signal Mfg Inc assumes no responsibility or liability for any errors or inaccuracies that may appear in this documentation TekTone the TekTone logo Tek ALERT Tek BRIDGE Tek CALL Tek...

Страница 3: ...ter Stations 19 Connect Data Bus Wiring 20 Connect Master Bus Wiring 20 Networking Multiple NC120 CE Modules 20 Configuration Setup 21 Connect the LS450 Config Tool Programming Software 21 Using the C...

Страница 4: ...ord stations 1 4 jack call cord sta tions duty stations dome lights and more In addition to these items the Tek CARE120 system can be easily integrated with exist ing TekTone nurse call systems nurse...

Страница 5: ...acility The monitor can also modify its own annunciation settings and display system faults The NC403TS uses a 22 inch touchscreen and may be mounted using the CL234 wall mount bracket Refer to IL1030...

Страница 6: ...ert staff to pending Tek CARE120 system calls The duty station is installed in areas where call annunciation is needed but the full functionality of a master station is not Four LEDs communicate the t...

Страница 7: ...each room in the facility and using the points on that room controller to connect to stations installed in that room In this installation strategy stations are normally wired to their default points o...

Страница 8: ...CE Module and ensure reliable system per formance the Tek CARE120 Power Calculator will calculate the master load for the module The cal culate may be downloaded at http www tektone com pdf_files gen2...

Страница 9: ...Figure 1 NC120 CE Module Connections Copyright TekTone Sound and Signal Mfg Inc All Rights Reserved IL992 Tek CARE120 Installation Manual 9 Installation...

Страница 10: ...0 CE port descriptions On the front of the NC120 CE Module you will find a bank of DIP switches These DIP switches are used to configure various options on the CE module and to provide an address for...

Страница 11: ...Bus The Tek CARE120 data bus is a two wire serial communications bus There are four ports on the front of the module each capable of driving up to 1 000 feet of data bus and up to 16 LI122 Series room...

Страница 12: ...to room controller and connect the room controllers to the data bus in parallel using T taps Connect the positive wire in the data bus cable to the solid white wire pin 13 on the back of the room cont...

Страница 13: ...ns may be installed on each port for a total of six master stations per NC120 CE Module To install multiple masters on one master bus trunk line use CT400P5 splitters Maximum tap length off of the mai...

Страница 14: ...t 0A and port 0B must be unique but port 0A and port 1A may have addresses duplicated between them Keep detailed notes of the physical addresses and locations of room controllers installed in the faci...

Страница 15: ...h each group of connections making up one point Refer to the figures below for default connection points Note Room controllers sold prior to warranty number N6861530 April 2017 were sup plied with Red...

Страница 16: ...connected to each point Do not splice multiple stations together on the same point wiring One to One Installations In most installations one room controller will be used as a connection point for all...

Страница 17: ...possible connect the patient station to Point 1 Purple and Purple White wires This is the default connection point for patient stations in a single or dual bed installation SF123 Two Button Pull Cord...

Страница 18: ...m the SF125 LED panel to the positive terminal of the desired point on the room controller and connect the black wire to the negative terminal of the desired point on the room con troller Make sure th...

Страница 19: ...iod opening the switch at any point during the check in period will satisfy the check in requirement For more information about check in see Check In on page 30 Connect NC415AV Master Stations The two...

Страница 20: ...tomatically assigned names Once the sys tem is powered on for the first time go to each master station on the system and touch the Menu button on the home screen then touch the This Master button that...

Страница 21: ...are the points on the LI122 Series and PM120 Room Controllers The points are the six physical connections on the rear of each controller shown as points 1 6 in the configuration In addition to the ph...

Страница 22: ...tions and or SF122 Dual Patient Stations per room or in an area where two or more Code Blue or Emergency switches are connected to the same room controller This scenario will be referred to as a Custo...

Страница 23: ...Figure 13 Config Tool Home Page Copyright TekTone Sound and Signal Mfg Inc All Rights Reserved IL992 Tek CARE120 Installation Manual 23 Configuration Setup...

Страница 24: ...enu Remove Private Data allows all private patient data such as medications con ditions etc to be stripped from the configuration so that it can be sent to TekTone for troubleshooting help without rev...

Страница 25: ...used to interact with the Tek CARE software These buttons appear at the top of editing panes in the Config Tool Edit Before information in an editing pane may be changed click this button to unlock th...

Страница 26: ...criptive file name and save the configuration Enabling and Naming Modules Navigate to the Modules page In the Modules list verify that all of the installed NC120 CE Modules installed on your system ar...

Страница 27: ...tation name For example a room controller addressed as 12 connected to port 0A or 0B of CE 01 would be assigned a default station name of 0112 If the same room controller address 12 were connected to...

Страница 28: ...edit multiple window is open select the Station IDs to be edited Once stations are selected enable the points that have stations connected and choose the correct station type and behavior Mark any unu...

Страница 29: ...lls from the stations in the selec ted zone Click Apply to save your changes Repeat for all zones required by the facility Room Level Reset SF123s can be programmed as Reset Buttons to enable the room...

Страница 30: ...which the system expects to see a check in input triggered If the check in input is not triggered during the scheduled time period an Inactivity call annunciates on the system Create several check in...

Страница 31: ...he Config Tool In the list of masters that appears select the master to be renamed Click the Edit button to unlock the Details pane for editing In the details pane give each master a unique descriptiv...

Страница 32: ...s are connected to Refer to the completed programming worksheet that was created during system installation to ensure cor rect programming The most important concept to remember during system configur...

Страница 33: ...ns Move Points to Stations Occasionally it may be necessary to assign a point to a different station than it is defaulted to To move points between Station IDs select the point to be moved from the po...

Страница 34: ...avior to a Station ID click Stations in the page selection pane Select the Station ID that the new hardware behavior is to be assigned to and click the Edit button in the Details pane Select the hardw...

Страница 35: ...editing the dome light annunciation for the Emergency call type Note that the settings shown above would result in a medium flash rate on the red LED of LI122s zoned to display the Emergency call howe...

Страница 36: ...the master sta tion If this option is available on your system the gray buttons on the right side of the NC415AV hous ing are used The buttons are labeled accordingly Using the NC415AV Master Station...

Страница 37: ...en displays information about the alarming station the type of call the resident name and more Staff requests Stat Attend and Status can also be placed from the details screen by touching the associat...

Страница 38: ...sired station will appear To enable or disable check in for a station dial the station touch the Check in button on the bottom of the screen Select any available check in schedules to apply to the sta...

Страница 39: ...ation including the Device ID of the current NC415AV the mod ule and port it is connected to the DIP switch address of the NC415AV and the Tek CARE software version Touch the Clean button to temporari...

Страница 40: ...Using the Tek CARE120 System Tek CARE120 Programming Worksheets 40 IL992 Tek CARE120 Installation Manual Copyright TekTone Sound and Signal Mfg Inc All Rights Reserved...

Отзывы: