45
Displayed alarms and their solutions
ALARMS
CAUSES AND SOLUTIONS
WHO CAN WORK
No or not enough water pressure to
compressor condenser.
Verify whether the water cock is open or
pressure is insufficient.
Break or malfunctioning of pressure valve. Call
REFRIGERATION TECHNICIAN
USER
Alarm only for MANTE 30 100
Probable gas leak or break of Low pressure
pressure switch.
Call Refrigeration Technician
REFRIGERATION
TECHNICIAN
Intervention of overload cut-out on Low Speed
of stirrer motor.
Call the Electrician to re-engage the overload
cut-out and verify its electric absorption.
ELECTRICIAN
Intervention of the overload cut out on High
Speed.
of the stirrer motor.
Call the Electrician to re-engage the overload
cut-out and verify its electric absorption.
ELECTRICIAN
1. Machine equipped with inverter.
Inverter block for overheating or fault.
If it is overheating wait and then re-start.
If it is faulty, call the Electrician.
2. If the machine is not equipped with
inverter, this alarm indicates that the
motor rotates in the wrong direction
and the phases are reversed.
ELECTRICIAN
Intervention of the overload cut out of the
refrigerator compressor
Call the Electrician to re-engage the overload
cut-out and verify its electric absorption.
Warns the front flange is open:
Close the flange and press ENTER to reset the
alarm
.
If the alarm persists it means the safety sensor
does not work and must be replaced.