Proline Promass F 500 EtherNet/IP
Operation options
93
3. Enter the password: serial number of the measuring device ex-works (e.g.
L100A802000).
LED on display module flashes: it is now possible to operate the measuring device
with the Web browser, FieldCare or DeviceCare.
The serial number can be found on the nameplate.
To ensure the safe and swift assignment of the WLAN network to the measuring
point, it is advisable to change the SSID name. It should be possible to clearly assign
the new SSID name to the measuring point (e.g. tag name) because it is displayed as
the WLAN network.
Disconnecting
‣
After configuring the device:
Terminate the WLAN connection between the operating unit and measuring device.
Starting the Web browser
1. Start the Web browser on the computer.
2. Enter the IP address of the Web server in the address line of the Web browser:
192.168.1.212
The login page appears.
6
7
8
9
10
1
5
2
3
4
A0029417
1
Picture of device
2
Device name
3
Device tag
4
Status signal
5
Current measured values
6
Operating language
7
User role
8
Access code
9
Login
10 Reset access code (→ 188)
If a login page does not appear, or if the page is incomplete → 211