System integration
Proline Promass F 500 EtherNet/IP
114
Description (format)
Byte
Bit
Value
28. External reference density unit (integer)
29-30
16
• 32840 (240): kg/Nm
3
• 32841 (240): kg/Nl
• 32842 (240): g/Scm
3
• 32843 (240): kg/Scm
3
• 32844 (240): lb/Sft
3
29. Not used
31-32
16
–
30. External temperature (real)
33-36
32
Data format:
Byte 1-4: External temperature
Floating-point number (IEEE754)
31. External temperature unit (integer)
37-38
16
• 1001 (32): °C
• 1002 (33): °F
• 1000 (35): K
• 1003 (34): °R
32. Not used
39-40
16
–
33. External value % S&W (real)
41-44
32
Data format:
Byte 1-4: External value, % S&W
Floating-point number (IEEE754)
34. External value, water cut % (real)
45-48
32
Data format:
Byte 1-4: External value, water cut %
Floating-point number (IEEE754)
35 Flow override monitoring
49-50
16
• 33004 (0): Off
• 33006 (1): On
36 Zero point adjustment monitoring
51-52
16
• 32823 (0): Cancel
• 33242 (0): Active
• 248 (0): Error zero point adjustment
• 33158 (1): Start
Permanently assigned configuration group
Config assembly (Assem104), 2704 byte
Description (format)
Bits
Byte
Offset
1.
None
32
4
0
2.
Parameter 36
–
Write protection
8
1
4
3.
None
8
1
5
4.
Parameter 87
System units
Mass flow unit
16
2
6
5.
Parameter 86
System units
Mass unit
16
2
8
6.
Parameter 93
System units
Volume flow unit
16
2
10
7.
Parameter 92
System units
Volume unit
16
2
12
8.
Parameter 80
System units
Corrected volume flow unit
16
2
14
9.
Parameter 79
System units
Corrected volume unit
16
2
16
10.
Parameter 81
System units
Density unit
16
2
18
11.
Parameter 89
System units
Reference density unit
16
2
20
12.
Parameter 91
System units
Temperature unit
16
2
22
13.
None
16
2
24
14.
Parameter 88
System units
Press. eng. unit
16
2
26
15.
Parameter 85
System units
Kinematic viscosity unit
16
2
28