Operation
Proline Promass F 500 EtherNet/IP
202
Output values for pulse/frequency/switch output
The
Pulse/frequency/switch output 1 to n
submenu contains all the parameters needed
to display the current measured values for every pulse/frequency/switch output.
Navigation
"Diagnostics" menu → Measured values → Output values → Pulse/frequency/switch output
1 to n
‣
Pulse/frequency/switch output
1 to n
Output frequency 1 to n
Pulse output 1 to n
Switch status 1 to n
Parameter overview with brief description
Parameter
Prerequisite
Description
User interface
Output frequency 1 to n
In the
Operating mode
parameter, the
Frequency
option is selected.
Displays the value currently measured
for the frequency output.
0.0 to 12
500.0 Hz
Pulse output 1 to n
In the
Operating mode
parameter, the
Pulse
option is selected.
Displays the pulse frequency currently
output.
Positive floating-point
number
Switch status 1 to n
The
Switch
option is selected in the
Operating mode
parameter.
Displays the current switch output
status.
• Open
• Closed
Output values for relay output
The
Relay output 1 to n
submenu contains all the parameters needed to display the
current measured values for every relay output.
Navigation
"Diagnostics" menu → Measured values → Output values → Relay output 1 to n
‣
Relay output 1 to n
Switch status
Switch cycles
Max. switch cycles number